11.040.70 - Ophthalmic equipment
ICS 11.040.70 Details
Ophthalmic equipment
Augenmedizinische Gerate
Matériel ophtalmique
Oftalmološka oprema
General Information
e-Library Subscription
Create subscription and get permanent access to documents within 11.040.70 - Ophthalmic equipment
Currently subscription includes documents marked with .We are working on making all documents available within the subscription.
ISO 19979:2018 provides guidance to manufacturers for the development of information to be provided to eye care practitioners for the hygienic management of trial hydrogel, composite and rigid gas-permeable (RGP) contact lenses intended for multipatient use.
ISO 19979:2018 does not apply to:
- labelling of contact lenses;
- the inactivation of prions and viruses since there are no standardised methods available for contact lenses.
ISO 19979:2018 can be used as guidance for the development of a hygienic management procedure for multipatient use.
NOTE ISO 14729 does not cover multipatient use.
- Standard16 pagesEnglish languagesale 10% offe-Library read for1 day
- Draft13 pagesEnglish languagesale 10% offe-Library read for1 day
This document specifies requirements for reference lenses for the calibration and verification of focimeters that are used for the measurement of spectacle form lenses, e.g. those complying with ISO 8598-1. It also gives a method for the determination of the back vertex power of the reference lenses.
NOTE It is accepted that other reference lenses can also be used with powers within the given range, manufactured to the same standard of accuracy and form, but different back vertex powers. However, only lenses with integer nominal powers, as described in 4.1, can be used for the calibration of digitally-rounding focimeters.
- Draft21 pagesEnglish languagesale 10% offe-Library read for1 day
This document specifies requirements for reference lenses for the calibration and verification of focimeters that are used for the measurement of spectacle form lenses, e.g. those complying with ISO 8598-1. It also gives a method for the determination of the back vertex power of the reference lenses. NOTE It is accepted that other reference lenses can also be used with powers within the given range, manufactured to the same standard of accuracy and form, but different back vertex powers. However, only lenses with integer nominal powers, as described in 4.1, can be used for the calibration of digitally-rounding focimeters.
- Standard19 pagesEnglish languagesale 15% off
- Standard19 pagesFrench languagesale 15% off
- Draft18 pagesEnglish languagesale 15% off
- Draft20 pagesFrench languagesale 15% off
ISO 19979:2018 provides guidance to manufacturers for the development of information to be provided to eye care practitioners for the hygienic management of trial hydrogel, composite and rigid gas-permeable (RGP) contact lenses intended for multipatient use.
ISO 19979:2018 does not apply to:
- labelling of contact lenses;
- the inactivation of prions and viruses since there are no standardised methods available for contact lenses.
ISO 19979:2018 can be used as guidance for the development of a hygienic management procedure for multipatient use.
NOTE ISO 14729 does not cover multipatient use.
- Standard16 pagesEnglish languagesale 10% offe-Library read for1 day
- Draft13 pagesEnglish languagesale 10% offe-Library read for1 day
This document, together with ISO 15004-1 and ISO 15004-2, specifies minimum requirements and test methods for hand-held, spectacle-type, and head-worn indirect ophthalmoscopes for observing indirect images of the eye fundus.
This document takes precedence over ISO 15004-1 and ISO 15004-2, if differences exist.
This document is not applicable to condensing lenses used for indirect ophthalmoscopy or to accessories.
This document is not applicable to table-mounted instruments such as Gullstrand ophthalmoscopes and their derivatives, nor to ophthalmoscopes primarily intended for image capture and/or processing such as those based on scanning laser techniques.
- Standard12 pagesEnglish languagesale 10% offe-Library read for1 day
- Draft8 pagesEnglish languagesale 10% offe-Library read for1 day
This document specifies requirements for the transmittance properties of uncut and unmounted finished spectacle lenses, including attenuation of solar radiation.
This document is not applicable to
— spectacle lenses having specific transmittance or absorption characteristics prescribed for medical reasons,
— products to which specific personal protective equipment transmittance standards apply, and
— products intended for direct observation of the sun, such as for solar-eclipse viewing.
NOTE 1 By reference to ISO 21987 and ISO 14889, this document also applies to lenses mounted in spectacles.
NOTE 2 Optical and geometric requirements are given for uncut finished spectacle lenses in ISO 8980‑1 and ISO 8980‑2, and for mounted lenses, in ISO 21987.
- Standard42 pagesEnglish languagesale 10% offe-Library read for1 day
- Draft37 pagesEnglish languagesale 10% offe-Library read for1 day
This document, together with ISO 15004-1 and ISO 15004-2, specifies minimum requirements and test methods for hand-held, spectacle-type, and head-worn indirect ophthalmoscopes for observing indirect images of the eye fundus.
This document takes precedence over ISO 15004-1 and ISO 15004-2, if differences exist.
This document is not applicable to condensing lenses used for indirect ophthalmoscopy or to accessories.
This document is not applicable to table-mounted instruments such as Gullstrand ophthalmoscopes and their derivatives, nor to ophthalmoscopes primarily intended for image capture and/or processing such as those based on scanning laser techniques.
- Standard12 pagesEnglish languagesale 10% offe-Library read for1 day
- Draft8 pagesEnglish languagesale 10% offe-Library read for1 day
This document, together with ISO 15004-1 and ISO 15004-2, specifies minimum requirements and test methods for hand-held, spectacle-type, and head-worn indirect ophthalmoscopes for observing indirect images of the eye fundus. This document takes precedence over ISO 15004-1 and ISO 15004-2, if differences exist. This document is not applicable to condensing lenses used for indirect ophthalmoscopy or to accessories. This document is not applicable to table-mounted instruments such as Gullstrand ophthalmoscopes and their derivatives, nor to ophthalmoscopes primarily intended for image capture and/or processing such as those based on scanning laser techniques.
- Standard5 pagesEnglish languagesale 15% off
- Standard5 pagesFrench languagesale 15% off
This document is applicable to ophthalmic viscosurgical devices (OVDs), a class of surgical invasive medical devices with viscous and/or viscoelastic properties, intended for use during surgery in the anterior segment of the human eye. OVDs are designed to create and maintain space, to protect intraocular tissues and to manipulate tissues during surgery.
This document specifies requirements with regard to safety for the intended performance, design attributes, preclinical and clinical evaluation, sterilization, product packaging, product labelling and information supplied by the manufacturer of these devices.
- Standard27 pagesEnglish languagesale 10% offe-Library read for1 day
- Draft22 pagesEnglish languagesale 10% offe-Library read for1 day
This document, together with ISO 15004-1 and ISO 15004-2, specifies minimum requirements and test methods for hand-held direct ophthalmoscopes designed for directly observing the eye fundus.
This document takes precedence over ISO 15004-1 and ISO 15004-2, if differences exist.
- Standard13 pagesEnglish languagesale 10% offe-Library read for1 day
- Draft11 pagesEnglish languagesale 10% offe-Library read for1 day
This document is applicable to all afocal (plano power) sunglasses and clip-ons for general use, including
road use and driving, intended for protection against solar radiation.
Information on the use of sunglass filters is given in Annex A. Requirements for unmounted filters used
as replacement or alternative filters are given in Annex C.
This document is not applicable to:
a) eyewear for protection against radiation from artificial light sources;
b) eye protectors intended for specific sports (e.g. ski goggles or other types – see ISO18527 (all parts));
c) sunglasses that have been medically prescribed for attenuating solar radiation;
d) products intended for direct observation of the sun, such as for viewing a partial or annular solar
eclipse, for which ISO 12312-2 applies;
e) products intended for occupational eye protection – see, for example, ISO 16321 (all parts)
- Standard40 pagesEnglish languagesale 10% offe-Library read for1 day
- Draft39 pagesEnglish languagesale 10% offe-Library read for1 day
IEC 80601-2-58:2014 applies to the basic safety and essential performance of lens removal devices and vitrectomy devices for ophthalmic surgery and associated accessories that can be connected to this medical electrical equipment, hereafter referred to as ME equipment. Hazards inherent in the intended physiological function of ME equipment or ME systems within the scope of this standard are not covered by specific requirements in this standard except in 7.2.13 and 8.4.1 of the general standard. This second edition includes changes in order to take into account the comments submitted during the approval of the first edition as a European Medical Device Directive, as well as the comments from other National Committees during the finalization of the first edition of this standard.
- Amendment14 pagesEnglish languagesale 10% offe-Library read for1 day
This document specifies requirements for the transmittance properties of uncut and unmounted
finished spectacle lenses, including attenuation of solar radiation.
This document is not applicable to
— spectacle lenses having specific transmittance or absorption characteristics prescribed for medical
reasons,
— products to which specific personal protective equipment transmittance standards apply, and
— products intended for direct observation of the sun, such as for solar-eclipse viewing.
NOTE 1 By reference to ISO 21987 and ISO 14889, this document also applies to lenses mounted in spectacles.
NOTE 2 Optical and geometric requirements are given for uncut finished spectacle lenses in ISO 8980-1
and ISO 8980-2, and for mounted lenses, in ISO 21987.
- Standard42 pagesEnglish languagesale 10% offe-Library read for1 day
- Draft37 pagesEnglish languagesale 10% offe-Library read for1 day
This document specifies requirements for the transmittance properties of uncut and unmounted finished spectacle lenses, including attenuation of solar radiation. This document is not applicable to — spectacle lenses having specific transmittance or absorption characteristics prescribed for medical reasons, — products to which specific personal protective equipment transmittance standards apply, and — products intended for direct observation of the sun, such as for solar-eclipse viewing. NOTE 1 By reference to ISO 21987 and ISO 14889, this document also applies to lenses mounted in spectacles. NOTE 2 Optical and geometric requirements are given for uncut finished spectacle lenses in ISO 8980‑1 and ISO 8980‑2, and for mounted lenses, in ISO 21987.
- Standard34 pagesEnglish languagesale 15% off
- Standard35 pagesFrench languagesale 15% off
This document applies to ocular endotamponades (OE), a group of non-solid surgically invasive medical devices introduced into the vitreous cavity of the eye to flatten and position a detached retina onto the retinal pigment epithelium (RPE), or to tamponade the retina.
With regard to the safety and efficacy of OE, this document specifies requirements for their intended performance, design attributes, pre-clinical and clinical evaluation, sterilization, product packaging, product labelling and the information supplied by the manufacturer.
- Standard23 pagesEnglish languagesale 10% offe-Library read for1 day
- Draft22 pagesEnglish languagesale 10% offe-Library read for1 day
This document, together with ISO 15004-1 and ISO 15004-2, specifies minimum requirements and test
methods for hand-held direct ophthalmoscopes designed for directly observing the eye fundus.
This document takes precedence over ISO 15004-1 and ISO 15004-2, if differences exist.
- Standard13 pagesEnglish languagesale 10% offe-Library read for1 day
- Draft11 pagesEnglish languagesale 10% offe-Library read for1 day
This document is applicable to ophthalmic viscosurgical devices (OVDs), a class of surgical invasive
medical devices with viscous and/or viscoelastic properties, intended for use during surgery in
the anterior segment of the human eye. OVDs are designed to create and maintain space, to protect
intraocular tissues and to manipulate tissues during surgery.
This document specifies requirements with regard to safety for the intended performance, design
attributes, preclinical and clinical evaluation, sterilization, product packaging, product labelling and
information supplied by the manufacturer of these devices.
- Standard27 pagesEnglish languagesale 10% offe-Library read for1 day
- Draft22 pagesEnglish languagesale 10% offe-Library read for1 day
This document is applicable to optical and electro-optical devices specified by the manufacturer for use by visually impaired persons as low vision aids. This document specifies requirements and test methods for optical and electro-optical devices specified by the manufacturer for use by visually impaired persons as low vision devices.
Implantable low vision devices are excluded.
- Standard35 pagesEnglish languagesale 10% offe-Library read for1 day
- Draft30 pagesEnglish languagesale 10% offe-Library read for1 day
This document, together with ISO 15004-1 and ISO 15004-2, specifies minimum requirements and test methods for hand-held direct ophthalmoscopes designed for directly observing the eye fundus. This document takes precedence over ISO 15004-1 and ISO 15004-2, if differences exist.
- Standard6 pagesEnglish languagesale 15% off
- Standard7 pagesFrench languagesale 15% off
- Draft6 pagesEnglish languagesale 15% off
- Draft6 pagesFrench languagesale 15% off
This document is applicable to ophthalmic viscosurgical devices (OVDs), a class of surgical invasive medical devices with viscous and/or viscoelastic properties, intended for use during surgery in the anterior segment of the human eye. OVDs are designed to create and maintain space, to protect intraocular tissues and to manipulate tissues during surgery. This document specifies requirements with regard to safety for the intended performance, design attributes, preclinical and clinical evaluation, sterilization, product packaging, product labelling and information supplied by the manufacturer of these devices.
- Standard19 pagesEnglish languagesale 15% off
- Standard20 pagesFrench languagesale 15% off
- Draft19 pagesEnglish languagesale 15% off
- Draft22 pagesFrench languagesale 15% off
This document specifies minimum requirements for instruments and systems that fall into the class of corneal topographers (CTs). It also specifies tests and procedures to verify that a system or instrument complies with this document and thus qualifies as a CT according to this document. It also specifies tests and procedures that allow the verification of capabilities of systems that are beyond the minimum requirements for CTs.
This document defines terms that are specific to the characterization of the corneal shape so that they may be standardized throughout the field of vision care.
This document is applicable to instruments, systems and methods that are intended to measure the surface shape of the cornea of the human eye.
NOTE The measurements can be of the curvature of the surface in local areas, three-dimensional topographical measurements of the surface or other more global parameters used to characterize the surface.
This document is not applicable to ophthalmic instruments classified as ophthalmometers.
- Standard29 pagesEnglish languagesale 10% offe-Library read for1 day
- Draft26 pagesEnglish languagesale 10% offe-Library read for1 day
This document applies to ocular endotamponades (OE), a group of non-solid surgically invasive medical
devices introduced into the vitreous cavity of the eye to flatten and position a detached retina onto the
retinal pigment epithelium (RPE), or to tamponade the retina.
With regard to the safety and efficacy of OE, this document specifies requirements for their intended
performance, design attributes, pre-clinical and clinical evaluation, sterilization, product packaging,
product labelling and the information supplied by the manufacturer.
- Standard23 pagesEnglish languagesale 10% offe-Library read for1 day
- Draft22 pagesEnglish languagesale 10% offe-Library read for1 day
This document is applicable to optical and electro-optical devices specified by the manufacturer for
use by visually impaired persons as low vision aids. This document specifies requirements and test
methods for optical and electro-optical devices specified by the manufacturer for use by visually
impaired persons as low vision devices.
Implantable low vision devices are excluded.
- Standard35 pagesEnglish languagesale 10% offe-Library read for1 day
- Draft30 pagesEnglish languagesale 10% offe-Library read for1 day
This document specifies minimum requirements for instruments and systems that fall into the class of
corneal topographers (CTs). It also specifies tests and procedures to verify that a system or instrument
complies with this document and thus qualifies as a CT according to this document. It also specifies
tests and procedures that allow the verification of capabilities of systems that are beyond the minimum
requirements for CTs.
This document defines terms that are specific to the characterization of the corneal shape so that they
may be standardized throughout the field of vision care.
This document is applicable to instruments, systems and methods that are intended to measure the
surface shape of the cornea of the human eye.
NOTE The measurements can be of the curvature of the surface in local areas, three-dimensional
topographical measurements of the surface or other more global parameters used to characterize the surface.
This document is not applicable to ophthalmic instruments classified as ophthalmometers.
- Standard29 pagesEnglish languagesale 10% offe-Library read for1 day
- Draft26 pagesEnglish languagesale 10% offe-Library read for1 day
This document is applicable to optical and electro-optical devices specified by the manufacturer for use by visually impaired persons as low vision aids. This document specifies requirements and test methods for optical and electro-optical devices specified by the manufacturer for use by visually impaired persons as low vision devices. Implantable low vision devices are excluded.
- Standard26 pagesEnglish languagesale 15% off
- Standard27 pagesFrench languagesale 15% off
- Draft25 pagesEnglish languagesale 15% off
- Draft27 pagesFrench languagesale 15% off
This document specifies minimum requirements for instruments and systems that fall into the class of corneal topographers (CTs). It also specifies tests and procedures to verify that a system or instrument complies with this document and thus qualifies as a CT according to this document. It also specifies tests and procedures that allow the verification of capabilities of systems that are beyond the minimum requirements for CTs. This document defines terms that are specific to the characterization of the corneal shape so that they may be standardized throughout the field of vision care. This document is applicable to instruments, systems and methods that are intended to measure the surface shape of the cornea of the human eye. NOTE The measurements can be of the curvature of the surface in local areas, three-dimensional topographical measurements of the surface or other more global parameters used to characterize the surface. This document is not applicable to ophthalmic instruments classified as ophthalmometers.
- Standard22 pagesEnglish languagesale 15% off
- Standard22 pagesFrench languagesale 15% off
- Draft22 pagesEnglish languagesale 15% off
- Draft22 pagesFrench languagesale 15% off
This document specifies fundamental requirements for non-invasive, active and non-active ophthalmic instruments and to devices for enhancing low vision. This document is also applicable to tonometers, but not to other ophthalmic instruments which are used in contact with the globe of the eye.
This document is not applicable to operation microscopes, endoscopes and devices intended for laser investigation or laser treatment of the eye.
- Standard14 pagesEnglish languagesale 10% offe-Library read for1 day
- Draft13 pagesEnglish languagesale 10% offe-Library read for1 day
- Amendment10 pagesEnglish languagesale 10% offe-Library read for1 day
- Draft7 pagesEnglish languagesale 10% offe-Library read for1 day
This document specifies particular requirements for the biocompatibility evaluation of materials for intraocular lenses (IOLs) including the processing conditions to produce them. These requirements include evaluation of physicochemical properties that are relevant to biocompatibility. It also gives guidance on conducting an ocular implantation test.
- Standard36 pagesEnglish languagesale 10% offe-Library read for1 day
- Draft33 pagesEnglish languagesale 10% offe-Library read for1 day
This document specifies a measuring system for spectacle frames and related vocabulary. It is applicable to spectacle frames with fronts that are intended to be symmetrical.
- Standard23 pagesEnglish languagesale 10% offe-Library read for1 day
- Draft19 pagesEnglish languagesale 10% offe-Library read for1 day
This document specifies fundamental requirements for non-invasive, active and non-active ophthalmic instruments and to devices for enhancing low vision. This document is also applicable to tonometers,
but not to other ophthalmic instruments which are used in contact with the globe of the eye.
This document is not applicable to operation microscopes, endoscopes and devices intended for laser investigation or laser treatment of the eye.
- Standard14 pagesEnglish languagesale 10% offe-Library read for1 day
- Draft13 pagesEnglish languagesale 10% offe-Library read for1 day
This document specifies particular requirements for the biocompatibility evaluation of materials for
intraocular lenses (IOLs) including the processing conditions to produce them. These requirements
include evaluation of physicochemical properties that are relevant to biocompatibility. It also gives
guidance on conducting an ocular implantation test.
- Standard36 pagesEnglish languagesale 10% offe-Library read for1 day
- Draft33 pagesEnglish languagesale 10% offe-Library read for1 day
- Amendment10 pagesEnglish languagesale 10% offe-Library read for1 day
- Draft7 pagesEnglish languagesale 10% offe-Library read for1 day
- Standard4 pagesEnglish languagesale 15% off
- Standard5 pagesFrench languagesale 15% off
This document specifies particular requirements for the biocompatibility evaluation of materials for intraocular lenses (IOLs) including the processing conditions to produce them. These requirements include evaluation of physicochemical properties that are relevant to biocompatibility. It also gives guidance on conducting an ocular implantation test.
- Standard27 pagesEnglish languagesale 15% off
- Standard28 pagesFrench languagesale 15% off
- Amendment7 pagesEnglish languagesale 10% offe-Library read for1 day
- Draft4 pagesEnglish languagesale 10% offe-Library read for1 day
This document specifies a measuring system for spectacle frames and related vocabulary. It is applicable
to spectacle frames with fronts that are intended to be symmetrical.
- Standard23 pagesEnglish languagesale 10% offe-Library read for1 day
- Draft19 pagesEnglish languagesale 10% offe-Library read for1 day
2019-12-04 - not allocated for publication because of missing TC decision to skip FV.
- Amendment7 pagesEnglish languagesale 10% offe-Library read for1 day
- Draft4 pagesEnglish languagesale 10% offe-Library read for1 day
This document applies to ocular endotamponades (OE), a group of non-solid surgically invasive medical devices introduced into the vitreous cavity of the eye to flatten and position a detached retina onto the retinal pigment epithelium (RPE), or to tamponade the retina. With regard to the safety and efficacy of OE, this document specifies requirements for their intended performance, design attributes, pre-clinical and clinical evaluation, sterilization, product packaging, product labelling and the information supplied by the manufacturer.
- Standard16 pagesEnglish languagesale 15% off
- Standard16 pagesEnglish languagesale 15% off
- Standard18 pagesFrench languagesale 15% off
This document specifies a measuring system for spectacle frames and related vocabulary. It is applicable to spectacle frames with fronts that are intended to be symmetrical.
- Standard16 pagesEnglish languagesale 15% off
- Standard16 pagesFrench languagesale 15% off
- Draft16 pagesEnglish languagesale 15% off
This document specifies fundamental requirements for non-invasive, active and non-active ophthalmic instruments and to devices for enhancing low vision. This document is also applicable to tonometers, but not to other ophthalmic instruments which are used in contact with the globe of the eye. This document is not applicable to operation microscopes, endoscopes and devices intended for laser investigation or laser treatment of the eye.
- Standard7 pagesEnglish languagesale 15% off
- Standard7 pagesFrench languagesale 15% off
- Amendment7 pagesEnglish languagesale 10% offe-Library read for1 day
- Draft4 pagesEnglish languagesale 10% offe-Library read for1 day
2019-12-04 - not allocated for publication because of missing TC decision to skip FV.
- Amendment7 pagesEnglish languagesale 10% offe-Library read for1 day
- Draft4 pagesEnglish languagesale 10% offe-Library read for1 day
- Standard1 pageEnglish languagesale 15% off
- Standard1 pageFrench languagesale 15% off
- Standard1 pageEnglish languagesale 15% off
- Standard1 pageFrench languagesale 15% off
This document defines terms relating to ophthalmic optics, specifically to blanks, finished spectacle lenses and fitting purposes.
Terms relating to processes and material for fabrication and surface treatment (other than some specific terms relating to coatings), and terms relating to defects in materials and after optical processing are given in ISO 9802.
- Standard76 pagesEnglish languagesale 10% offe-Library read for1 day
This document describes the most commonly used test methods considered in standardization work relating to scratch and abrasion resistance of plastic spectacle lenses along with their technical capacities and limitations. It includes the ISO test method for assessment of claims for basic abrasion resistance in ISO 8980-5. This document is intended to be of benefit to any future interest in ISO standardization on scratch and abrasion resistance of spectacle lenses.
- Technical report19 pagesEnglish languagesale 15% off
- Technical report21 pagesFrench languagesale 15% off
IEC 80601-2-58:2014 applies to the basic safety and essential performance of lens removal devices and vitrectomy devices for ophthalmic surgery and associated accessories that can be connected to this medical electrical equipment, hereafter referred to as ME equipment. Hazards inherent in the intended physiological function of ME equipment or ME systems within the scope of this standard are not covered by specific requirements in this standard except in 7.2.13 and 8.4.1 of the general standard. This second edition includes changes in order to take into account the comments submitted during the approval of the first edition as a European Medical Device Directive, as well as the comments from other National Committees during the finalization of the first edition of this standard.
- Amendment14 pagesEnglish languagesale 10% offe-Library read for1 day
This document defines terms applicable to intraocular lenses, and to the methods used to evaluate them.
NOTE Terms are listed in the alphabetical order of the English terms in the English version of this document.
- Standard17 pagesEnglish languagesale 10% offe-Library read for1 day
This document defines terms relating to ophthalmic optics, specifically to blanks, finished spectacle
lenses and fitting purposes.
Terms relating to processes and material for fabrication and surface treatment (other than some specific
terms relating to coatings), and terms relating to defects in materials and after optical processing are
given in ISO 9802.
- Standard76 pagesEnglish languagesale 10% offe-Library read for1 day