Nuclear fuel technology — Controlled-potential coulometric assay of plutonium

ISO 12183:2005 specifies an analytical method for the electrochemical assay of pure plutonium nitrate solutions of nuclear grade, with a total uncertainty of 0,1 % to 0,2 % at the confidence level of 0,95 for a single determination. The method is suitable for aqueous solutions containing more than 0,5 g/L plutonium and test samples containing between 4 mg and 15 mg of plutonium. Application of this technique to solutions containing less than 0,5 g/L, and test samples containing less than 4 mg of plutonium, must be demonstrated by the user as having adequate reliability for their specific application.

Technologie du combustible nucléaire — Dosage du plutonium par coulométrie à potentiel imposé

L'ISO 12183:2005 spécifie une méthode analytique pour le dosage électrochimique de plutonium pur dans des solutions nitriques de qualité nucléaire, avec une incertitude totale de 0,1 % à 0,2 % sous un intervalle de confiance de 0,95 pour une simple détermination. La méthode est adaptée aux solutions aqueuses contenant plus de 0,5 g/L de plutonium et des échantillons contenant entre 4 mg et 15 mg de plutonium. L'application de cette technique aux solutions contenant moins de 0,5 g/L et d'échantillons contenant moins de 4 mg de plutonium doit être démontrée par l'utilisateur comme ayant une fiabilité adéquate pour ces applications spécifiques.

General Information

Status
Withdrawn
Publication Date
07-Feb-2005
Withdrawal Date
07-Feb-2005
Current Stage
9599 - Withdrawal of International Standard
Completion Date
05-Aug-2016
Ref Project

Relations

Buy Standard

Standard
ISO 12183:2005 - Nuclear fuel technology -- Controlled-potential coulometric assay of plutonium
English language
24 pages
sale 15% off
Preview
sale 15% off
Preview
Standard
ISO 12183:2005 - Technologie du combustible nucléaire -- Dosage du plutonium par coulométrie a potentiel imposé
French language
26 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)

INTERNATIONAL ISO
STANDARD 12183
Second edition
2005-02-01

Nuclear fuel technology — Controlled-
potential coulometric assay of plutonium
Technologie du combustible nucléaire — Dosage du plutonium par
coulométrie à potentiel imposé




Reference number
ISO 12183:2005(E)
©
ISO 2005

---------------------- Page: 1 ----------------------
ISO 12183:2005(E)
PDF disclaimer
This PDF file may contain embedded typefaces. In accordance with Adobe's licensing policy, this file may be printed or viewed but
shall not be edited unless the typefaces which are embedded are licensed to and installed on the computer performing the editing. In
downloading this file, parties accept therein the responsibility of not infringing Adobe's licensing policy. The ISO Central Secretariat
accepts no liability in this area.
Adobe is a trademark of Adobe Systems Incorporated.
Details of the software products used to create this PDF file can be found in the General Info relative to the file; the PDF-creation
parameters were optimized for printing. Every care has been taken to ensure that the file is suitable for use by ISO member bodies. In
the unlikely event that a problem relating to it is found, please inform the Central Secretariat at the address given below.


©  ISO 2005
All rights reserved. Unless otherwise specified, no part of this publication may be reproduced or utilized in any form or by any means,
electronic or mechanical, including photocopying and microfilm, without permission in writing from either ISO at the address below or
ISO's member body in the country of the requester.
ISO copyright office
Case postale 56 • CH-1211 Geneva 20
Tel. + 41 22 749 01 11
Fax + 41 22 749 09 47
E-mail copyright@iso.org
Web www.iso.org
Published in Switzerland

ii © ISO 2005 – All rights reserved

---------------------- Page: 2 ----------------------
ISO 12183:2005(E)
Contents Page
Foreword. iv
1 Scope. 1
2 Normative references . 1
3 Principle . 1
4 Reagents . 2
5 Apparatus. 2
6 Procedure. 7
6.1 Plutonium determination. 7
6.2 Analysis of subsequent samples . 12
7 Expression of results. 12
7.1 Calculation of the electrical calibration factor. 12
7.2 Calculation of the blank. 12
7.3 Fraction of electrolysed plutonium . 13
7.4 Plutonium content. 14
7.5 Quality control. 14
8 Characteristics of the method . 14
8.1 Repeatability. 14
8.2 Confidence interval. 14
8.3 Analysis time . 15
9 Interferences. 15
10 Procedure variations and optimisation . 17
10.1 Accountability measurements. 17
10.2 Process control measurements . 18
10.3 Measurement cell design . 18
10.4 Electrolyte and electrode options . 18
10.5 Test sample size. 19
10.6 Background current corrections . 19
10.7 Correction for iron . 19
10.8 Control-potential adjustment. 20
Annex A (normative) Purification by anion-exchange separation . 21
Annex B (normative) Determination of Formal Potential E . 23
0
Bibliography . 24

© ISO 2005 – All rights reserved iii

---------------------- Page: 3 ----------------------
ISO 12183:2005(E)
Foreword
ISO (the International Organization for Standardization) is a worldwide federation of national standards bodies
(ISO member bodies). The work of preparing International Standards is normally carried out through ISO
technical committees. Each member body interested in a subject for which a technical committee has been
established has the right to be represented on that committee. International organizations, governmental and
non-governmental, in liaison with ISO, also take part in the work. ISO collaborates closely with the
International Electrotechnical Commission (IEC) on all matters of electrotechnical standardization.
International Standards are drafted in accordance with the rules given in the ISO/IEC Directives, Part 2.
The main task of technical committees is to prepare International Standards. Draft International Standards
adopted by the technical committees are circulated to the member bodies for voting. Publication as an
International Standard requires approval by at least 75 % of the member bodies casting a vote.
Attention is drawn to the possibility that some of the elements of this document may be the subject of patent
rights. ISO shall not be held responsible for identifying any or all such patent rights.
ISO 12183 was prepared by Technical Committee ISO/TC 85, Nuclear energy, Subcommittee SC 5, Nuclear
fuel technology.
This second edition cancels and replaces the first edition (ISO 12183:1995), which has been technically
revised.

iv © ISO 2005 – All rights reserved

---------------------- Page: 4 ----------------------
INTERNATIONAL STANDARD ISO 12183:2005(E)

Nuclear fuel technology — Controlled-potential coulometric
assay of plutonium
1 Scope
This International Standard specifies an analytical method for the electrochemical assay of pure plutonium
nitrate solutions of nuclear grade, with a total uncertainty of 0,1 % to 0,2 % at the confidence level of 0,95 for a
single determination. The method is suitable for aqueous solutions containing more than 0,5 g/L plutonium
and test samples containing between 4 mg and 15 mg of plutonium. Application of this technique to solutions
containing less than 0,5 g/L, and test samples containing less than 4 mg of plutonium, must be demonstrated
by the user as having adequate reliability for their specific application.
Preliminary purification by anion exchange is required to remove interfering substances when present in
significant amounts. Purification is also appropriate in situations where the purity of the sample is unknown or
when it may unpredictably fluctuate in the manufacturing process. Refer to Clause 9 for a discussion of
interferences and iron corrections.
Clause 10 discusses the changes in application of the method and methodology that can be applied and
important considerations when selecting measurement parameters, while still remaining within the intended
scope of this International Standard.
2 Normative references
The following referenced documents are indispensable for the application of this document. For dated
references, only the edition cited applies. For undated references, the latest edition of the referenced
document (including any amendments) applies.
ISO 10980:1995, Validation of the strength of reference solutions used for measuring concentrations
3 Principle
The method consists of the following steps:
 test samples are prepared by weight and fumed with sulfuric acid to achieve a consistent and stable
chemical form that is free from chloride, fluoride, nitrate, nitrite, hydroxylamine, and volatile organic
compounds;
 when appropriate, preliminary purification by anion exchange with fuming in sulfuric acid to restore the dry
plutonium sulfate chemical form in preparation for measurement;
 measuring a blank of the nitric-acid supporting electrolyte and performing an appropriate calculation to
determine the background current correction applicable to the electrolysis of the test sample from
charging, faradic, and residual current; [1]
 dissolution of the residue in the previously measured supporting electrolyte (the blank);
© ISO 2005 – All rights reserved 1

---------------------- Page: 5 ----------------------
ISO 12183:2005(E)
3+
 reduction at a controlled potential that electrolyses the plutonium to greater than 99,8 % Pu and
measurement of the equilibrium solution potential at the end of this electrolysis by control-potential
adjustment; [2]
4+
 oxidation at a controlled potential that electrolyses the plutonium to greater than 99,8 % Pu and
measurement of the equilibrium solution potential at the end of this electrolysis by control-potential
adjustment;
 periodically, the formal potential of the plutonium couple, E is measured for the specific supporting
0,
electrolyte, cell components, and design;
 the result is corrected for the background current and the fraction of plutonium not electrolysed;
 coulometer calibration, traceable to electrical standards through Ohm’s law and Faraday’s constant, is
used to convert the coulombs of integrated current from the electrolyses to equivalents of plutonium;
 quality-control plutonium standards are used to independently demonstrate measurement performance.
4 Reagents
Use only reagents of recognized analytical grade.
All aqueous solutions shall be prepared with double-distilled or deionized water with a resistivity greater than
10 MΩ⋅cm.
4.1 Nitric acid solution, c(HNO ) = 0,9 mol/L.
3
4.2 Amidosulfuric acid solution, c(NH HSO ) = 1,5 mol/L.
2 3
4.3 Sulfuric acid solution, c(H SO ) = 3 mol/L.
2 4
4.4 Pure argon or nitrogen, (O content lower than 10 parts per million).
2
4.5 Pure air, free of any organic compounds.
5 Apparatus
Usual laboratory equipment found in a medium-activity radiochemical laboratory suitable for work with
plutonium.
5.1 Analytical balance, capable of weighing after installation in radiological containment with an accuracy
of 0,1 mg, or less. (The test sample size must be correspondingly greater than the minimum 1 g size, if the
accuracy, as installed, does not meet this 0,1 mg criteria.)
5.2 Weighing burette, glass or plastic; the material selection is not critical provided it maintains a stable
tare weight and is not susceptible to static charge.
5.3 Equipment for sample evaporation in the coulometric cell, comprising the following:
 an overhead radiant heater or hotplate with a means of temperature adjustment;
 for optimum evaporation rate and to minimize the concentration of acid fumes, an air supply with the
delivery tube directed towards the surface of the liquid may be used;
2 © ISO 2005 – All rights reserved

---------------------- Page: 6 ----------------------
ISO 12183:2005(E)
 additional design feature to control acid fumes, including vapour capture and local neutralization may be
appropriate depending upon facility design and requirements.
The evaporation/fuming apparatus shall be adjusted to provide both rapid and well-controlled evaporation,
followed by fuming to dryness of the plutonium sulfate test sample. The apparatus shall prevent mechanical
loss of the test-sample solution from boiling and spattering during the process and from contamination by
chemicals, which may be used to neutralize acid vapours.
The equipment in Figure 1 is suitable.
Dimensions in centimetres

Figure 1 — Sample evaporation system
5.4 Controlled-potential coulometer.
The following equipment is suitable (see Figure 2).
5.4.1 Coulometer-cell assembly, comprising the following:
−1
 a stirrer motor with a rotation frequency of at least 1 000 min ;
NOTE Adjustable-speed motors allow optimizing for individual cell designs. Stirring motors powered by isolated
direct-current supplies have the advantage of reducing electrical noise that is superimposed on the blank and test-sample
electrolysis current signal sent to the integrator.
 a cylindrical or tapered glass coulometric cell of capacity 50 mL, or less, with an O-ring seal and a lid
suitable for the following internal equipment:
 an inlet tube for inert gas, to exclude atmospheric oxygen;
© ISO 2005 – All rights reserved 3

---------------------- Page: 7 ----------------------
ISO 12183:2005(E)
 a chemically inert stirrer, designed to prevent splashing, centrally located, and connected to a stirrer
motor;
 a working electrode made of gold (99,99 %) and consisting of a gold wire welded or machined to a
cylindrical gold wire frame, nominally 15 mm high and 20 mm in diameter, around which is welded or
machined a very fine gold mesh. The gold mesh may be of several layers (e.g, four layers);
 a glass tube plugged at the bottom end with a 2,5 mm thick sintered-glass disc (pore size < 0,01 µm),
filled with nitric acid (4.1) and the tip positioned within the ring of the working electrode;
 a reference electrode, saturated calomel electrode (SCE), or other reference electrodes as described
in 10.3, housed in the glass tube;
 another glass tube, similar to the first one, also filled with nitric acid (4.1), and mounted within the ring
of the working electrode;
 an auxiliary electrode consisting of a platinum wire, 0,5 mm to 3,0 mm in diameter, housed in the
glass tube;
The diameter of the glass tube and sintered-glass disc used to house the auxiliary electrode may be larger
than the glass tube used for the reference electrode. The flow rate of the solution through both glass discs
shall be less than 0,05 mL/h.
a) A thermocouple for measuring the temperature of the test-sample solution is an optional feature. If it is
not possible to measure the solution temperature either during or immediately following the analysis, then
a thermometer inside the containment unit may be used to estimate the solution temperature from the
ambient air temperature. An accuracy of ± 0,1 °C is recommended.
b) For optimum potential control, design the cell assembly to position the sintered-glass discs of the glass
tubes to meet the following requirements:
 the reference electrode connection is 1 mm to 2 mm from the working electrode;
 the distance between auxiliary and reference electrode discs is less than or equal to the distance
between the auxiliary electrode disc and the nearest point on the working electrode.
c) The hole for the stirrer shaft provides a gas vent. Except for the stirrer passage, all openings in the lid
shall be relatively gastight such that the inert-gas flow rate selected is sufficient to quickly degas oxygen
from the supporting electrolyte and the test-sample solution and to prevent in-leakage of atmospheric
oxygen. For ease of operation and consistent performance, the gas-flow inlet tube may be designed to
direct all or part of the inert gas toward the test-sample solution, such that a 2 mm to 4 mm dimple is
produced on the sample solution without causing the solution to splash. For cell assemblies with an
optimized design and fit, degassing is accomplished in 5 min or less.
4 © ISO 2005 – All rights reserved

---------------------- Page: 8 ----------------------
ISO 12183:2005(E)

1 video 8 auxiliary (or counter) electrode in bridge tube filled with supporting
electrolyte
2 printer
9 reference electrode in bridge tube filled with supporting electrolyte
3 control computer
10 inert gas
4 keyboard
11 stirrer
5 potentiostat and integrator
12 working electrode
6 voltmeter digital
7 AC or DC power for stirring motor 13 cell
Figure 2 — Coulometric-cell assembly connections
© ISO 2005 – All rights reserved 5

---------------------- Page: 9 ----------------------
ISO 12183:2005(E)
5.4.2 Measuring unit, comprising the following [3,4].
a) A potentiostat with the desired range of electrolysis potentials for plutonium measurement and the
following capabilities:
 a high power, minimum current output capability of 250 mA;
 a quick-response control-potential circuit, maximum rise-time from zero volts to the desired control
potential of 1 ms, with maximum voltage overshoot of 1 mV;
 a control amplifier with a common-mode rejection of 90 dB or greater;
 automatic control-potential adjustment, with an output stepping capability resolution of 0,001 V, or
less;
 a voltage-follower amplifier used to isolate the reference electrode (electrometer), with a minimum
11
input impedance of 10 Ω;
 detection capability for monitoring the full range of the electrolysis currents, including a sensitivity of
0,5 µA at zero current.
NOTE This procedure assumes that the coulometer has two accurate potentiometers for selecting the oxidation and
reduction potentials, although this is not a system requirement.
b) A coulometric integrator capable of being read to ± 10 µC and able to integrate currents varying from
1 µA to 150 mA. (Refer to 6.1.4 for integrator capabilities and calibration requirements.)
 The control-potential system should not drift more than ± 1 mV and the integration system should not
drift more than 0,005 % during routine measurements (between electrical calibrations), over the
range of temperatures to which the control-potential circuitry will be exposed. The system should be
located in a cabinet having adequate temperature control to limit electronic drift within the specified
accuracy, if the room temperature varies excessively.
 An electronic clock, accurate within the selected range to better than 0,002 %.
 A system for generating a known constant current, stable to 0,002 % over the range of temperatures
to which the constant-current circuitry will be exposed. This system will be used for electrical
calibration of the integration circuit of the coulometer, as described in 6.1.4.
 The cable connecting the potentiostat to the cell should be a three-wire conductor, twisted-shielded
cable, preferably with the shield grounded at the potentiostat. Gold connectors at the cell are
recommended to avoid contact problems from corrosion.
5.5 Digital voltmeter (DVM), having an accuracy of ± 0,001 % of the voltage reading in the range 0,5 V to
10
10 V, and 0,01 % of the voltage reading in the range 100 mV to 500 mV, with an input impedance of 10 Ω or
greater. This accuracy is required for electrical calibration, as described in 6.1.4.
5.6 Regulated power: equipment should be protected with an uninterruptable power supply that provides a
regulated voltage of ± 1 % of the national standard, and provides appropriate surge-protection features.
6 © ISO 2005 – All rights reserved

---------------------- Page: 10 ----------------------
ISO 12183:2005(E)
6 Procedure
6.1 Plutonium determination
6.1.1 Weighing the test sample, with an accuracy of 0,01 %.
The test sample may be weighed after delivery into a tared coulometer cell and the apparent mass corrected
for the air buoyancy effect using Equation (1).
Alternatively, the test sample may be delivered by weighing into the coulometer cell, as follows:
NOTE Care is needed to avoid any loss of microdrops. During weighing of the polyethylene weighing burette, the
problem of static electricity is eliminated by contact between the dropping tube and a copper plate connected to the
ground or a similar arrangement.
For test samples with concentrations greater than 15 g/L of plutonium, the solution should be diluted by weight
so that all weights are greater than 1 g and the overall weighing uncertainty remains acceptably small.
a) Fill a polyethylene weighing burette with the solution to be analysed.
b) Weigh the burette to 0,1 mg.
c) Deliver a test sample of at least 1 mL, drop by drop, into a coulometric cell.
d) Weigh the burette again to 0,1 mg.
e) The mass difference gives the apparent mass, m , of the test sample in the cell.
a
f) Correct the apparent mass of the test sample for the air buoyancy affect using Equation (1):
1−DD
( )
ab
mm=× (1)
Real a
1− D D
()
as
where
D is the density of air, which is a function of temperature, pressure, and humidity, but typically is
a
3 3
0,001 16 g/cm to 0,001 2 g/cm ;
3
D is the density of the stainless steel weights used in modern analytical balances, 8,0 g/cm ;
b
D is the density of the test sample.
s
In addition to applying an air buoyancy correction to the mass of the test sample, air buoyancy corrections
should be applied to all mass measurements (including any bulk-material mass measurements). This
correction is required to eliminate systematic errors that can exceed 0,1 % for solutions. The correction is less
for solids, but can still be significant.
6.1.2 Preparation of the test sample
a) Add 1 mL of sulfuric acid solution (4.3) to the coulometric cell containing the test sample.
b) Place the cell containing the test sample into the sample evaporation system and evaporate the liquid in
the test sample.
c) Evaporate at a temperature sufficient to evolve SO fumes, and continue until SO fumes are no longer
3 3
observed and a residue of plutonium sulfate (orange-coloured precipitate) is formed. Do not allow the
solution to boil or splash as this will cause mechanical loss.
© ISO 2005 – All rights reserved 7

---------------------- Page: 11 ----------------------
ISO 12183:2005(E)
The colour of the plutonium sulfate is dependent on the type of lighting used in the laboratory. Under
fluorescent lighting, the dried sulfate is coral pink. Degradation of plutonium sulfate to plutonium oxide should
not be expected, even after baking the residue, unless extreme temperatures are employed. Failure to use
high-purity reagents, anion-exchange resins washed free of resin fines, and heating equipment that is not well
maintained and clean can produce visible black residue in combination with the dried sulfate powder. These
residues could be mistaken for plutonium oxide, and, depending on their composition, might interfere in
subsequent measurements.
d) Allow the test sample to cool to room temperature.
e) If the presence of significant amounts of impurities is suspected, purify the plutonium in the test sample
by a suitable purification process, then repeating the sulfuric acid fuming step as detailed in 6.1.2. Anion-
exchange separation as outlined in Annex A is an effective purification process.
6.1.3 Electrode pretreatment
Electrode conditioning is critical and shall be repeated periodically. The laboratory shall therefore have the
capability to perform aggressive electrode treatments. The following techniques may be used individually, or in
combination with each other, to treat the working and auxiliary electrodes:
 storing in 8 mol/L nitric acid (recommended as the general overnight practice);
 soaking in concentrated sulfuric acid containing 10 % hydrofluoric acid, followed by 8 mol/L nitric acid;
 boiling in nitric acid;
 soaking in aqua regia (limited to several minutes to prevent damage to the working electrode);
 flaming the platinum auxiliary electrode to white or red heat.
Electrode treatment may be performed on a preventative basis, at the beginning and/or the end of the day that
the electrode is used. Alternatively, treatment may be on an “as needed” basis, particularly in the case of
failure to obtain optimum electrode performance when measuring either the blank or the test sample. The
typical background current levels should be relatively constant for a given installation and are normally used
as indicators of the desired performance.
Each day, or more often as desired, before performing the actual blank determination, further conditioning of
the electrodes is achieved by performing the following sequence of electrolyses.
a) Assemble the cell lid complete with the electrodes and other internal equipment (see 5.4.1).
b) Take a clean dry coulometric cell and add sufficient nitric acid solution (4.1) to immerse the working
electrode, and the sintered-glass tips of the reference and auxiliary electrode tubes.
c) Add one drop of amidosulfuric acid solution (4.2).
d) Firmly fit the cell under the lid.
e) Start the stirrer at the desired speed. (Using the maximum speed, while avoiding splashing or forming any
excessive vortex that would interrupt electrical connections, is recommended.)
f) Open the gas inlet and maintain a sufficient flow of inert gas throughout the electrolysis period.
(Inadequate purging to remove oxygen can be mistaken for an electrode-conditioning problem.)
g) Preselect the oxidation potential at E + 0,32 V and the reduction potential at E − 0,36 V.
0 0
h) After degassing for 5 min, start the oxidation and oxidize at E + 0,32 V until a residual current of 10 µA is
0
obtained.
8 © ISO 2005 – All rights reserved

---------------------- Page: 12 ----------------------
ISO 12183:2005(E)
i) Start the reduction and reduce at E − 0,36 V until a residual current lower than 10 µA is obtained.
0
j) Oxidize at E + 0,32 V.
0
k) Stop the electrolysis when the current is lower than 10 µA.
l) Based upon electrode performance:
 proceed with the blank determination, see 6.1.6.c); or
 rinse the electrolysis cell and the outside wall of the fritted-glass tubes with fresh supporting
electrolyte to prepare it for further conditioning, see 6.1.3; or
 rinse the electrolysis cell and the outside wall of the fritted-glass tubes with fresh supporting
electrolyte to prepare it for blank determination, see 6.1.6.
6.1.4 Electrical calibration of the current integration system
The electrical calibration factor of the coulometer is measured by using a high-accuracy, highly stable
constant current in place of the electrolysis cell. Detailed instructions for the calibration of a current integration
system are highly dependent upon the design of the specific integration circuit. However, the following
principles and specifications apply to achieve an accuracy of ∼ 0,01 % for this aspect of the overall
measurement process:
Generate a constant current (stable and accurately known to within ∼ 0,002 %) in a manner that is
electronically equivalent to the process by which the electrolysis current from the test sample and the blank
are integrated. (Typically, the potentiostat is converted into a constant current source by replacing the cell
assembly with a standard resistor. The voltage drop across the standard resistor is accurately measured to
accurately determine the actual calibration current. Alternatively, voltage and resistance measurements are
used to periodically calibrate an external constant current source.)
Time the duration of calibration within ∼ 0,002 %.
The linearity of the integrator response shall be demonstrated for the range of currents observed during
plutonium measurement (∼ 50 µA − 100 mA). The impact on plutonium measurements from integrator
nonlinearity shall be less than 0,005 %, to avoid introducing excessive uncertainty in the electrical calibration
process.
A typical sequence for performing an electrical calibration is the following:
a) configure the instrumentation for electrical calibration and set to the desired constant current, for example
10,000 mA;
b) set the integration time to an
...

NORME ISO
INTERNATIONALE 12183
Deuxième édition
2005-02-01


Technologie du combustible nucléaire —
Dosage du plutonium par coulométrie à
potentiel imposé
Nuclear fuel technology — Controlled-potential coulometric assay of
plutonium




Numéro de référence
ISO 12183:2005(F)
©
ISO 2005

---------------------- Page: 1 ----------------------
ISO 12183:2005(F)
PDF – Exonération de responsabilité
Le présent fichier PDF peut contenir des polices de caractères intégrées. Conformément aux conditions de licence d'Adobe, ce fichier
peut être imprimé ou visualisé, mais ne doit pas être modifié à moins que l'ordinateur employé à cet effet ne bénéficie d'une licence
autorisant l'utilisation de ces polices et que celles-ci y soient installées. Lors du téléchargement de ce fichier, les parties concernées
acceptent de fait la responsabilité de ne pas enfreindre les conditions de licence d'Adobe. Le Secrétariat central de l'ISO décline toute
responsabilité en la matière.
Adobe est une marque déposée d'Adobe Systems Incorporated.
Les détails relatifs aux produits logiciels utilisés pour la création du présent fichier PDF sont disponibles dans la rubrique General Info
du fichier; les paramètres de création PDF ont été optimisés pour l'impression. Toutes les mesures ont été prises pour garantir
l'exploitation de ce fichier par les comités membres de l'ISO. Dans le cas peu probable où surviendrait un problème d'utilisation,
veuillez en informer le Secrétariat central à l'adresse donnée ci-dessous.


©  ISO 2005
Droits de reproduction réservés. Sauf prescription différente, aucune partie de cette publication ne peut être reproduite ni utilisée sous
quelque forme que ce soit et par aucun procédé, électronique ou mécanique, y compris la photocopie et les microfilms, sans l'accord écrit
de l'ISO à l'adresse ci-après ou du comité membre de l'ISO dans le pays du demandeur.
ISO copyright office
Case postale 56 • CH-1211 Geneva 20
Tel. + 41 22 749 01 11
Fax. + 41 22 749 09 47
E-mail copyright@iso.org
Web www.iso.org
Publié en Suisse

ii © ISO 2005 – Tous droits réservés

---------------------- Page: 2 ----------------------
ISO 12183:2005(F)
Sommaire Page
Avant-propos. iv
1 Domaine d'application. 1
2 Références normatives. 1
3 Principe . 1
4 Réactifs . 2
5 Appareillage. 2
6 Mise en œuvre . 7
6.1 Mesure de plutonium. 7
6.2 Analyse d’échantillons successifs. 12
7 Expression des résultats. 12
7.1 Calcul du facteur d’étalonnage électrique. 12
7.2 Calcul du blanc. 13
7.3 Fraction de plutonium électrolysé . 14
7.4 Plutonium contenu. 14
7.5 Contrôle qualité. 15
8 Caractéristiques de la méthode. 15
8.1 Répétitivité. 15
8.2 Intervalle de confiance . 15
8.3 Temps d’analyse . 15
9 Interférences. 15
10 Variantes et optimisation de la procédure .18
10.1 Mesures pour comptabilité . 18
10.2 Mesures de contrôle de procédé. 19
10.3 Conception de la cellule de mesure. 19
10.4 Options pour les électrolyte et électrode . 19
10.5 Taille de l’échantillon d’essai . 20
10.6 Corrections de courants de fond . 20
10.7 Correction du fer . 21
10.8 Ajustement du potentiel de contrôle. 21
Annexe A (normative) Purification par séparation par échange anionique . 23
Annexe B (normative) Détermination du potentiel formel, E . 25
0
Bibliographie . 26

© ISO 2005 – Tous droits réservés iii

---------------------- Page: 3 ----------------------
ISO 12183:2005(F)
Avant-propos
L'ISO (Organisation internationale de normalisation) est une fédération mondiale d'organismes nationaux de
normalisation (comités membres de l'ISO). L'élaboration des Normes internationales est en général confiée
aux comités techniques de l'ISO. Chaque comité membre intéressé par une étude a le droit de faire partie du
comité technique créé à cet effet. Les organisations internationales, gouvernementales et non
gouvernementales, en liaison avec l'ISO participent également aux travaux. L'ISO collabore étroitement avec
la Commission électrotechnique internationale (CEI) en ce qui concerne la normalisation électrotechnique.
Les Normes internationales sont rédigées conformément aux règles données dans les Directives ISO/CEI,
Partie 2.
Les projets de Normes internationales adoptés par les comités techniques sont soumis aux comités membres
pour vote. Leur publication comme Normes internationales requiert l'approbation de 75 % au moins des
comités membres votants.
L'attention est appelée sur le fait que certains des éléments du présent document peuvent faire l'objet de
droits de propriété intellectuelle ou de droits analogues. L'ISO ne saurait être tenue pour responsable de ne
pas avoir identifié de tels droits de propriété et averti de leur existence.
L’ISO 12183 a été élaborée par le comité technique ISO/TC 85, Énergie nucléaire, sous-comité SC 5,
Technologie du combustible nucléaire.
Cette deuxième édition annule et remplace la première édition (ISO 12183:1995) qui a fait l’objet d’une
révision technique
iv © ISO 2005 – Tous droits réservés

---------------------- Page: 4 ----------------------
NORME INTERNATIONALE ISO 12183:2005(F)

Technologie du combustible nucléaire — Dosage du plutonium
par coulométrie à potentiel imposé
1 Domaine d'application
La présente Norme internationale spécifie une méthode analytique pour le dosage électrochimique de
plutonium pur dans des solutions nitriques de qualité nucléaire, avec une incertitude totale de 0,1 % à 0,2 %
sous un intervalle de confiance de 0,95 pour une simple détermination. La méthode est adaptée aux solutions
aqueuses contenant plus de 0,5 g/l de plutonium et des échantillons contenant entre 4 mg et 15 mg de
plutonium. L'application de cette technique aux solutions contenant moins de 0,5 g/l et d'échantillons
contenant moins de 4 mg de plutonium doit être démontrée par l'utilisateur comme ayant une fiabilité
adéquate pour ces applications spécifiques.
Une purification préalable par échange anionique est nécessaire pour éliminer les substances interférentes
présentes en quantités significatives. La purification est aussi nécessaire dans les situations où la pureté de
l'échantillon n'est pas connue ou quand elle peut fluctuer de façon imprévisible dans le procédé de fabrication.
L’Article 9 traite des interférences et de la correction du fer.
L’Article 10 traite des changements dans l'application de la méthode et de la méthodologie qui peut être
appliquée et des considérations importantes pour sélectionner les paramètres de mesure, tout en restant à
l'intérieur du domaine d'application de la présente Norme internationale.
2 Références normatives
Les documents de référence suivants sont indispensables pour l'application du présent document. Pour les
références datées, seule l'édition citée s'applique. Pour les références non datées, la dernière édition du
document de référence s'applique (y compris les éventuels amendements).
ISO 10980:1995, Vérification du titre des solutions de référence utilisées pour la mesure des concentrations
3 Principe
La méthode comprend les étapes suivantes:
 les échantillons sont préparés par pesée et portés à sec en présence d'acide sulfurique pour obtenir une
forme solide et stable chimiquement, exempte de chlorure, fluorure, nitrate, nitrite, hydroxylamine et de
composés organiques volatils;
 si nécessaire, une purification préalable par échange ionique avec mise à sec dans de l'acide sulfurique
pour revenir à la forme chimique sèche de sulfate de plutonium prête pour la mesure;
 mesurer un blanc de l'électrolyte à base d'acide nitrique et réaliser les calculs appropriés pour déterminer
la correction de courant de fond à appliquer à l'électrolyse de l'échantillon d’essai due au courant de
charge, faradique et résiduel [1];
 dissolution du résidu dans l'électrolyte préalablement mesuré (le blanc);
© ISO 2005 – Tous droits réservés 1

---------------------- Page: 5 ----------------------
ISO 12183:2005(F)
3+
 réduction à potentiel contrôlé qui électrolyse le plutonium à plus de 99,8 % en Pu et mesure du
potentiel de la solution à l'équilibre à la fin de l'électrolyse par ajustement du potentiel de contrôle [2];
4+
 oxydation à potentiel contrôlé qui électrolyse le plutonium à plus de 99,8 % en Pu et mesure du
potentiel de la solution à l'équilibre à la fin de l'électrolyse par ajustement du potentiel de contrôle;
 périodiquement, le potentiel formel du couple plutonium, E , est mesuré pour un électrolyte, des
0
composants de cellule et une conception spécifiques;
 le résultat est corrigé du courant de base et de la fraction du plutonium qui n'est pas électrolysé;
 l'étalonnage du coulomètre, rattaché aux références électriques par la loi d'Ohm et la constante de
Faraday, est utilisé pour convertir les coulombs du courant intégré de l'électrolyse en équivalents
plutonium;
 des étalons plutonium de contrôle qualité sont utilisés pour démontrer de façon indépendante la qualité
des mesurages.
4 Réactifs
N'utiliser que des réactifs de qualité analytique reconnue.
Toutes les solutions aqueuses doivent être préparées avec de l'eau bidistillée ou désionisée ayant une
résistivité supérieure à 10 MΩ⋅cm.
4.1 Solution d'acide nitrique, c(HNO ) = 0,9 mol/l.
3
4.2 Solution d’acide amidosulfurique, c(NH HSO ) = 1,5 mol/l.
2 3
4.3 Solution d’acide sulfurique, c(H SO ) = 3 mol/l.
2 4
4.4 Argon ou azote pur, teneur en oxygène inférieure à 10 ppm.
4.5 Air pur, exempt de tous composés organiques.
5 Appareillage
Équipement courant de laboratoire de radiochimie de moyenne activité, adapté au travail avec du plutonium.
5.1 Balance analytique, capable de peser après installation en milieu confiné avec une précision d'au
moins 0,1 mg. (La taille de l'échantillon doit être supérieure au minimum de 1 g, si la précision, après
installation, n'atteint pas le critère de 0,1 mg.)
5.2 Burette à peser, en verre ou en matière plastique, le choix du matériau n'est pas critique à condition
qu'il assure une tare stable et ne soit pas sujet à l'électricité statique.
5.3 Équipement pour évaporation de l'échantillon dans la cellule coulométrique, comprenant les
éléments suivants:
 un chauffage radiant par-dessus ou une plaque chauffante avec réglage de température;
 une alimentation d'air avec un tube d'alimentation dirigé vers la surface du liquide peut être utilisée afin
d’obtenir une vitesse d'évaporation optimale et diminuer les vapeurs acides;
2 © ISO 2005 – Tous droits réservés

---------------------- Page: 6 ----------------------
ISO 12183:2005(F)
 des caractéristiques additionnelles pour contrôler les vapeurs acides, incluant le piégeage des vapeurs et
leur neutralisation localement peuvent être utiles selon l'installation et les nécessités.
L'appareil d'évaporation/mise à sec doit être réglé pour permettre à la fois une évaporation rapide et bien
contrôlée, suivie d’une mise à sec de l’échantillon de sulfate de plutonium. L'appareil doit empêcher la perte
mécanique de la solution d'échantillon par ébullition et éclaboussures pendant le processus et la
contamination par des produits chimiques qui peuvent être utilisés pour neutraliser les vapeurs acides.
L'équipement de la Figure 1 est adapté.
Dimensions en centimètres

Figure 1 — Système d'évaporation d'échantillon
5.4 Coulomètre à potentiel imposé
Le matériel présenté à la Figure 2 est adapté.
5.4.1 Ensemble de cellule de coulomètre, comprenant les éléments suivants:
−1
 un agitateur motorisé avec une fréquence de rotation supérieure à 1 000 min .
NOTE Des moteurs à vitesse ajustable permettent une optimisation de la conception des cellules individuelles. Les
moteurs d'agitateur alimentés par des courants continus isolés ont l'avantage de réduire le bruit électrique qui est
superposé au blanc et au courant d'électrolyse de l'échantillon envoyé à l'intégrateur.
 une cellule coulométrique en verre cylindrique ou effilée de 50 ml de capacité, ou moins, avec un joint
torique et un couvercle adapté pour les équipements internes suivants:
 un tube d'injection de gaz inerte pour éliminer l'oxygène atmosphérique;
© ISO 2005 – Tous droits réservés 3

---------------------- Page: 7 ----------------------
ISO 12183:2005(F)
 un agitateur inerte chimiquement, conçu pour éviter les éclaboussures, placé au centre, et connecté
à son moteur;
 une électrode de travail faite d'or (99,99 %) et comprenant un fil d'or soudé ou usiné à un cadre
cylindrique de fil d'or, 15 mm de haut et 20 mm de diamètre, autour duquel est soudée ou usinée une
très fine toile d'or, la toile d'or peut être en plusieurs couches (par exemple 4 couches);
 un tube de verre plongeant avec à son extrémité basse un disque fritté de 2,5 mm de diamètre (taille
de pore < 0,01 µm), rempli d'acide nitrique (4.1) et l'extrémité placée dans l'anneau de l'électrode de
travail;
 une électrode de référence, électrode au calomel saturé (ECS), ou autres électrodes de référence
comme décrit en 10.3, logée dans le tube de verre;
 un autre tube de verre, similaire au premier, rempli aussi d'acide nitrique (4.1), et monté dans
l'anneau de l'électrode de travail;
 une électrode auxiliaire comprenant un fil de platine, de 0,5 mm à 3,0 mm de diamètre, logée dans le
tube de verre.
Le diamètre du tube de verre et du disque fritté utilisé pour loger l'électrode auxiliaire peut être plus grand que
celui du tube utilisé pour l'électrode de référence. Le débit de la solution à travers l'ensemble des disques
frittés doit être inférieur à 0,05 ml/h.
a) Un thermocouple pour mesurer la température de la solution d'échantillon est optionnel. S'il n'est pas
possible de mesurer la température de la solution soit pendant soit immédiatement après l'analyse, alors
un thermomètre à l'intérieur de la cellule de confinement peut être utilisé pour estimer la température de
la solution à partir de la température de l'air ambiant. Une précision de ± 0,1 °C est recommandée.
b) Pour un contrôle optimal du potentiel, concevoir l'ensemble de la cellule afin de positionner les disques
frittés des tubes de verre en respectant les exigences suivantes:
 la jonction de l'électrode de référence est entre 1 mm et 2 mm de l'électrode de travail;
 la distance entre les disques des électrodes de référence et auxiliaire est inférieure ou égale à la
distance entre le disque de l'électrode auxiliaire et le plus proche point de l'électrode de travail.
c) Le trou de l'arbre de l'agitateur assure l'échappement des gaz. À l'exception du passage de l'agitateur,
toutes les ouvertures doivent être relativement étanches aux gaz de façon que le débit de gaz inerte
choisi soit suffisant pour dégazer rapidement l'oxygène dissous de l'électrolyte et de la solution
d'échantillon et pour éviter une fuite d'oxygène atmosphérique. Pour faciliter l'opération et l'uniformité des
performances, le tube d'injection du flux gazeux peut être conçu pour diriger tout ou partie du gaz inerte
vers la solution de l'échantillon, de telle façon qu'une ride de 2 mm à 4 mm soit produite sur la solution
d'échantillon sans provoquer d'éclaboussure. Pour les ensembles de cellule de conception optimisée et
ajustée, le dégazage est réalisé en 5 min ou moins.
4 © ISO 2005 – Tous droits réservés

---------------------- Page: 8 ----------------------
ISO 12183:2005(F)

Légende
1 vidéo 8 électrode auxiliaire en tube du pont électrique rempli d’électrolyte
2 imprimante 9 électrode de référence en tube du pont électrique rempli
d’électrolyte
3 ordinateur de contrôle
10 gaz inerte
4 clavier
11 agitateur
5 potentiostat et intégrateur
6 voltmètre numérique 12 électrode de travail
13 cellule
7 alimentation cc ou ca pour le moteur
de l’agitateur
Figure 2 — Schéma de l’ensemble de coulométrie
© ISO 2005 – Tous droits réservés 5

---------------------- Page: 9 ----------------------
ISO 12183:2005(F)
5.4.2 Unité de mesure, comprenant les éléments suivants [3, 4]:
a) un potentiostat avec une étendue de potentiels adaptée pour les mesures du plutonium et avec les
possibilités suivantes
 une grande puissance, courant minimal de sortie possible de 250 mA;
 un circuit de contrôle de potentiel à réponse rapide, avec un temps de montée de zéro au potentiel
contrôlé désiré de 1 ms et avec une tension de dépassement instantanée de 1 mV;
 un amplificateur de contrôle avec une réjection de mode commun de 90 dB ou plus;
 un ajustement de potentiel contrôlé, avec une résolution de pas de sortie de 0,001 V, ou moins;
 un amplificateur suiveur de tension utilisé pour isoler l'électrode de référence (électromètre), avec
11
une impédance d'entrée de 10 Ω minimale;
 une capacité de détection pour le contrôle de toute la gamme des courants d'électrolyse, incluant
une sensibilité de 0,5 µA à courant nul.
NOTE Ce mode opératoire suppose que le coulomètre a deux potentiomètres précis pour sélectionner les potentiels
d'oxydation et de réduction, bien que cela ne soit pas impératif.
b) un intégrateur coulométrique, pouvant être lu à ± 10 µC et capable d'intégrer des courants variant de
1 µA à 150 mA, (voir 6.1.4 pour les capacités de l'intégrateur et les besoins d'étalonnage).
 Le système de contrôle du potentiel ne doit pas dériver de plus de ± 1 mV et le système d’intégration
ne doit pas dériver de plus de 0,005 % pendant les mesures courantes (entre les étalonnages
électriques), sur toute la gamme de température à laquelle les circuits de contrôle de potentiel seront
exposés. Il convient de placer le système dans une armoire ayant un contrôle de température adapté
pour limiter la dérive électronique dans le domaine spécifié, si la température ambiante varie
fortement.
 Une horloge électronique, avec une précision supérieure à 0,002 % dans la gamme de temps choisie.
 Un système de génération de courant constant, stable à 0,002 % dans la gamme de température à
laquelle le circuit sera exposé. Ce système sera utilisé pour l’étalonnage électrique du circuit
d’intégration du coulomètre, comme décrit en 6.1.4.
 Il convient que le câble connectant le potentiostat à la cellule soit un triple conducteur, torsadé et
blindé, de préférence avec un blindage relié à la masse du potentiostat. Des connecteurs en or sur la
cellule sont recommandés pour éviter les problèmes de contact dus à la corrosion.
5.5 Voltmètre numérique, ayant une précision de ± 0,001 % de la valeur de voltage lue dans la gamme
allant de 0,5 V à 10 V, et de 0,01 % du voltage lu dans la gamme de 100 mV à 500 mV, avec une impédance
10
d’entrée de 10 Ω ou plus. Cette précision est nécessaire pour l’étalonnage électrique, comme indiqué en
6.1.4.
5.6 Alimentation régulée, il convient d'équiper l'équipement avec une source de courant permanent
délivrant une tension à ± 1 % du standard national, et disposant de caractéristiques de protection adaptées
contre les pics.
6 © ISO 2005 – Tous droits réservés

---------------------- Page: 10 ----------------------
ISO 12183:2005(F)
6 Mise en œuvre
6.1 Mesure de plutonium
6.1.1 Pesée de l’échantillon d’essai, avec une précision de ± 0,01 %.
L’échantillon d’essai peut être pesé après dépôt dans une cellule de coulométrie tarée et la masse apparente
corrigée de la poussée de l’air en utilisant l’Équation (1).
En variante, l’échantillon d’essai peut être déposé par pesée dans la cellule du coulomètre, comme suit:
NOTE Un soin particulier doit être apporté pour éviter toute perte de microgouttes. Pendant la pesée de la burette de
polyéthylène, le problème de l’électricité statique est éliminé en réalisant un contact entre le tube de distribution de goutte
et une plaque de cuivre connectée à la terre ou par un montage équivalent.
Pour les échantillons d’essai ayant des concentrations supérieures à 15 g/l de plutonium, il convient que la
solution subisse une dilution massique telle que toutes les masses soient supérieures à 1 g et que l’incertitude
totale de pesée reste suffisamment petite.
a) Remplir une burette à peser en polyéthylène avec la solution à analyser.
b) Peser la burette à 0,1 mg près.
c) Délivrer, goutte à goutte, au moins 1 ml de l’échantillon d’essai dans la cellule de coulométrie.
d) Peser encore la burette à 0,1 mg près.
e) La différence de masse donne la masse apparente, m , de l’échantillon d’essai dans la cellule.
a
f) Corriger la masse apparente de l’échantillon d’essai de la poussée de l’air en utilisant l’Équation (1):
1−DD
( )
ab
mm=× (1)
Réelle a
1−DD
()
as

D est la masse volumique de l’air, qui est fonction de la température, de la pression et de l’humidité,
a
3 3
mais pratiquement varie de 0,001 16 g/cm à 0,001 2 g/cm ;
D est la masse volumique de l’acier inoxydable des masses utilisées dans les balances analytiques
b
3
modernes, 8,0 g/cm ;
D est la masse volumique de l’échantillon d’essai.
s
En plus de l’application de la correction de poussée de l’air à la masse de l’échantillon d’essai, il convient que
la correction de poussée de l’air soit appliquée à toutes les mesures de masse (incluant toutes les mesures
de masse des matières en vrac). Ces corrections sont nécessaires pour éliminer les erreurs systématiques
qui peuvent excéder 0,1 % pour les solutions. La correction est plus faible pour les solides, mais peut être
encore significative.
6.1.2 Préparation de l’échantillon d’essai
a) Ajouter 1 ml d’acide sulfurique (4.3) à la cellule coulométrique contenant l’échantillon d’essai.
b) Placer la cellule contenant l’échantillon d’essai dans le système d’évaporation et évaporer le liquide de
l’échantillon d’essai.
© ISO 2005 – Tous droits réservés 7

---------------------- Page: 11 ----------------------
ISO 12183:2005(F)
c) Évaporer à une température suffisante pour former des fumées de SO et continuer jusqu’à ce que les
3
fumées de SO ne soient plus visibles et qu’un résidu de sulfate de plutonium (précipité de couleur
3
orange) soit formé. Ne pas faire bouillir la solution ou éclabousser, ce qui pourrait conduire à des pertes
mécaniques de matière.
La couleur du sulfate de plutonium dépend du type de lumière utilisée dans le laboratoire. Sous éclairage
fluorescent, le sulfate sec est rose corail. Des dégradations de sulfate de plutonium en oxyde de
plutonium ne devraient pas apparaître, même après cuisson du résidu, à moins que des températures
extrêmes n’aient été atteintes. La non-utilisation de réactifs de grande pureté, de résines échangeuses
d’anions lavées exemptes de particules fines de résine, et un équipement de chauffage qui n’est pas
correctement entretenu et propre peuvent produire des traces visibles noires dans la poudre sèche de
sulfate. Ces résidus peuvent être pris par erreur pour de l’oxyde de plutonium et, suivant leur composition,
peuvent interférer dans les mesurages ultérieurs.
d) Laisser l’échantillon d’essai refroidir à la température de la salle.
e) Si la présence de quantités significatives d’impuretés est suspectée, purifier le plutonium de l’échantillon
d’essai par un procédé de purification adapté, puis répéter l’étape de mise à sec en milieu sulfurique
détaillée dans le chapitre 6.1.2. La séparation par échange anionique présentée à l’Annexe A est un
procédé efficace de purification.
6.1.3 Traitement préliminaire de l’électrode
Le conditionnement de l’électrode est critique et doit être répété régulièrement. Le laboratoire doit donc avoir
la possibilité de réaliser des traitements agressifs de l’électrode. Les techniques suivantes peuvent être
utilisées individuellement ou combinées entre elles, pour traiter les électrodes de travail et auxiliaire:
 les stocker dans l’acide nitrique 8 mol/l (recommandé comme pratique générale pour la nuit);
 trempage dans l’acide sulfurique concentré contenant 10 % d’acide fluorhydrique, suivi d’acide nitrique
8 mol/l;
 faire bouillir dans l’acide nitrique;
 trempage dans l’eau régale (limité à quelques minutes pour éviter des dommages sur l’électrode de
travail);
 passer l’électrode auxiliaire en platine à la flamme jusqu’au rouge cerise ou blanc.
Le traitement de l’électrode peut être réalisé de façon préventive, au début et/ou à la fin de la journée de
l’utilisation de l’électrode. Alternativement, un traitement peut être fait chaque fois que nécessaire, en
particulier en cas de difficulté à obtenir une performance optimale de l’électrode quand on mesure le blanc ou
l’échantillon d’essai. Les niveaux de courant de fond habituels devraient être relativement constants pour une
installation donnée et sont normalement utilisés comme indicateurs de la performance désirée.
Chaque jour, ou plus si souhaité, avant de réaliser une mesure de blanc, d’autres conditionnement des
électrodes sont réalisés en appliquant les séquences d’électrolyse suivantes.
a) Assembler le couvercle de la cellule avec ses électrodes et ses autres équipements internes (voir 5.4.1).
b) Prendre une cellule coulométrique sèche et propre et ajouter suffisamment d’acide nitrique (4.1) pour
immerger l’électrode de travail, et les extrémités en verre fritté des tubes des électrodes de référence et
auxiliaire.
c) Ajouter une goutte de solution d’acide amidosulfurique (4.2).
d) Ajuster fermement la cellule sous son couvercle.
8 © ISO 2005 – Tous droits réservés

---------------------- Page: 12 ----------------------
ISO 12183:2005(F)
e) Démarrer l’agitateur à la vitesse désirée. (L’utilisation d’une vitesse maximum, tout en évitant les
éclaboussures ou la formation d’un vortex trop important qui interromprait les connections électriques, est
recommandée.)
f) Ouvrir l’injection de gaz et maintenir un débit suffisant de gaz inerte pendant toute la période de
l’électrolyse. (Une chasse inadaptée pour enlever l’oxygène pourrait être prise pour un défaut de
conditionnement d’électrode.)
g) Présélectionner les potentiels d’oxydation à E + 0,32 V et de réduction à E − 0,36 V.
0 0
h) Après un dégazage de 5 min, démarrer l’oxydation et oxyder à E + 0,32 V jusqu’à l’obtention d’un
0
courant résiduel de 10 µA.
i) Démarrer la réduction et réduire à E − 0,36 V jusqu’à obtenir un courant résiduel de 10 µA.
0
j) Oxyder à E + 0,32 V.
0
k) Arrêter l’électrolyse quand le courant est inférieur à 10 µA.
l) Selon les performances de
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.