Radiological protection — Minimum criteria for electron paramagnetic resonance (EPR) spectroscopy for retrospective dosimetry of ionizing radiation — Part 1: General principles

The primary purpose of ISO 13304-1:2013 is to provide minimum acceptable criteria required to establish procedure of retrospective dosimetry by electron paramagnetic resonance spectroscopy and to report the results. The second purpose is to facilitate the comparison of measurements related to absorbed dose estimation obtained in different laboratories. ISO 13304-1:2013 covers the determination of absorbed dose in the measured material. It does not cover the calculation of dose to organs or to the body. It covers measurements in both biological and inanimate samples, and specifically: based on inanimate environmental materials, usually made at X-band microwave frequencies (8 GHz to 12 GHz); in vitro tooth enamel using concentrated enamel in a sample tube, usually employing X-band frequency, but higher frequencies are also being considered; in vivo tooth dosimetry, currently using L-band (1 GHz to 2 GHz), but higher frequencies are also being considered; in vitro nail dosimetry using nail clippings measured principally at X-band, but higher frequencies are also being considered; in vivo nail dosimetry with the measurements made at X-band on the intact finger or toe; in vitro measurements of bone, usually employing X-band frequency, but higher frequencies are also being considered. For the biological samples, the in vitro measurements are carried out in samples after their removal from the person and under laboratory conditions, whereas the measurements in vivo may take place under field conditions. NOTE The dose referred to in ISO 13304-1:2013 is the absorbed dose of ionizing radiation in the measured materials.

Radioprotection — Critères minimaux pour la spectroscopie par résonance paramagnétique électronique (RPE) pour la dosimétrie rétrospective des rayonnements ionisants — Partie 1: Principes généraux

Le but principal de l'ISO 13304-1:2013 est de fournir un ensemble de critères minimaux acceptables requis pour établir une procédure de dosimétrie rétrospective par spectroscopie par résonance paramagnétique électronique et pour rendre compte des résultats. Le but secondaire de l'ISO 13304-1:2013 vise à faciliter la comparaison entre laboratoire des mesurages et de l'estimation de la dose absorbée. L'ISO 13304-1:2013 couvre la détermination de la dose absorbée dans le matériau mesuré. Elle ne couvre pas le calcul de la dose délivrée aux organes ou à l'organisme entier. Elle ne concerne que les mesurages effectuées sur des échantillons biologiques et des échantillons inertes, et plus particulièrement: a) les mesurages de matériaux environnementaux inertes, généralement réalisés des fréquences micro-ondes de la bande X (8 GHz à 12 GHz); b) les mesurages in vitro de prélèvement d'émail dentaire, placé dans un tube porte-échantillon, et mesuré en général en bande X, mais l'utilisation de fréquences micro-ondes plus élevées peut être également considérée; c) les mesurages in vivo de dents, réalisés actuellement en bande L (1 GHz à 2 GHz), mais des fréquences micro-ondes plus élevées sont également envisagées; d) les mesurages in vitro de prélèvements d'ongles effectués principalement dans la bande X, mais des fréquences micro-ondes plus élevées sont également à l'étude; e) les mesurages in vivo des ongles, effectués en bande X sur les ongles des doigts ou des orteils; f) les mesurages in vitro de tissus osseux, réalisés en général en bande X mais l'utilisation de fréquences micro-ondes plus élevées est également étudiée. En ce qui concerne les échantillons biologiques, les mesurages in vitro sont effectués sur des échantillons prélevés sur la personne et dans des conditions de laboratoire, tandis que les systèmes de mesure in vivo, réalisés sur les individus peuvent être déplacés au plus près des victimes.

General Information

Status
Withdrawn
Publication Date
14-Jul-2013
Withdrawal Date
14-Jul-2013
Current Stage
9599 - Withdrawal of International Standard
Completion Date
20-Jul-2020
Ref Project

Relations

Buy Standard

Standard
ISO 13304-1:2013 - Radiological protection -- Minimum criteria for electron paramagnetic resonance (EPR) spectroscopy for retrospective dosimetry of ionizing radiation
English language
18 pages
sale 15% off
Preview
sale 15% off
Preview
Standard
ISO 13304-1:2013 - Radioprotection -- Criteres minimaux pour la spectroscopie par résonance paramagnétique électronique (RPE) pour la dosimétrie rétrospective des rayonnements ionisants
French language
20 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)

INTERNATIONAL ISO
STANDARD 13304-1
First edition
2013-07-01
Radiological protection — Minimum
criteria for electron paramagnetic
resonance (EPR) spectroscopy for
retrospective dosimetry of ionizing
radiation —
Part 1:
General principles
Radioprotection — Critères minimaux pour la spectroscopie par
résonance paramagnétique électronique (RPE) pour la dosimétrie
rétrospective des rayonnements ionisants —
Partie 1: Principes généraux
Reference number
ISO 13304-1:2013(E)
©
ISO 2013

---------------------- Page: 1 ----------------------
ISO 13304-1:2013(E)

COPYRIGHT PROTECTED DOCUMENT
© ISO 2013
All rights reserved. Unless otherwise specified, no part of this publication may be reproduced or utilized otherwise in any form
or by any means, electronic or mechanical, including photocopying, or posting on the internet or an intranet, without prior
written permission. Permission can be requested from either ISO at the address below or ISO’s member body in the country of
the requester.
ISO copyright office
Case postale 56 • CH-1211 Geneva 20
Tel. + 41 22 749 01 11
Fax + 41 22 749 09 47
E-mail copyright@iso.org
Web www.iso.org
Published in Switzerland
ii © ISO 2013 – All rights reserved

---------------------- Page: 2 ----------------------
ISO 13304-1:2013(E)

Contents Page
Foreword .iv
Introduction .v
1 Scope . 1
2 Terms and definitions . 1
3 Confidentiality and ethical considerations . 2
4 Laboratory safety requirements . 2
4.1 Magnetic field . 2
4.2 Electromagnetic frequency . 3
4.3 Biohazards from samples . 3
5 Collection/selection and identification of samples . 3
6 Transportation and storage of samples . 3
7 Preparation of samples. 4
8 Apparatus . 5
8.1 Principles of EPR spectroscopy . 5
8.2 Requirements for EPR spectrometers . 5
8.3 Requirements for the resonator . 5
8.4 Measurements of the background signals . 6
8.5 Spectrometer stability and monitoring/control of environmental conditions . 6
8.6 Baseline drift . 6
9 Measurements of the samples . 7
9.1 General principles . 7
9.2 Choice and optimization of the measurement parameters . 7
9.3 Sample positioning and loading . 9
9.4 Microwave bridge tuning . 9
9.5 Use of standard samples as field markers and amplitude monitors . 9
9.6 Monitoring reproducibility .10
9.7 Procedure to measure anisotropic samples .10
9.8 Coding of spectra and samples .10
10 Determination of the absorbed dose in the samples .10
10.1 Determination of the radiation-induced signal intensity .10
10.2 Conversion of the EPR signal into an estimate of absorbed dose .11
11 Measurement uncertainty .11
12 Investigation of dose that has been questioned .12
13 Quality assurance (QA) and quality control (QC) .13
14 Minimum documentation requirements .14
Bibliography .15
© ISO 2013 – All rights reserved iii

---------------------- Page: 3 ----------------------
ISO 13304-1:2013(E)

Foreword
ISO (the International Organization for Standardization) is a worldwide federation of national standards
bodies (ISO member bodies). The work of preparing International Standards is normally carried out
through ISO technical committees. Each member body interested in a subject for which a technical
committee has been established has the right to be represented on that committee. International
organizations, governmental and non-governmental, in liaison with ISO, also take part in the work.
ISO collaborates closely with the International Electrotechnical Commission (IEC) on all matters of
electrotechnical standardization.
The procedures used to develop this document and those intended for its further maintenance are
described in the ISO/IEC Directives, Part 1. In particular the different approval criteria needed for the
different types of ISO documents should be noted. This document was drafted in accordance with the
editorial rules of the ISO/IEC Directives, Part 2. www.iso.org/directives
Attention is drawn to the possibility that some of the elements of this document may be the subject of
patent rights. ISO shall not be held responsible for identifying any or all such patent rights. Details of
any patent rights identified during the development of the document will be in the Introduction and/or
on the ISO list of patent declarations received. www.iso.org/patents
Any trade name used in this document is information given for the convenience of users and does not
constitute an endorsement.
The committee responsible for this document is ISO/TC 85, Nuclear energy, nuclear technologies, and
radiological protection, Subcommittee SC 2, Radiological protection.
ISO 13304 consists of the following parts, under the general title Radiological protection — Minimum criteria
for electron paramagnetic resonance (EPR) spectroscopy for retrospective dosimetry of ionizing radiation:
— Part 1: General principles
iv © ISO 2013 – All rights reserved

---------------------- Page: 4 ----------------------
ISO 13304-1:2013(E)

Introduction
Electron paramagnetic resonance (EPR) has become an important approach for retrospective dosimetry
in any situation where dosimetric information is potentially incomplete or unknown for an individual.
It is now applied widely for retrospective evaluation of doses that were delivered at considerable times
in the past (e.g. EPR dosimetry is one of the methods of choice for retrospective evaluation of doses to
the involved populations from the atomic weapon exposures in Japan and after the Chernobyl accident)
and has received attention for use for triage after an incident in which large numbers of people have
potentially been exposed to clinically significant levels of radiation. Various materials may be analysed
by EPR to provide information on dose. Thus, EPR is a versatile tool for retrospective dosimetry,
pertinent as well for acute exposures (past or recent, whole or partial body) and prolonged exposures.
Doses estimated with EPR were mainly used to correlate the biological effect of ionizing radiation to
received dose, to validate other techniques or methodologies, to manage casualties, or for forensic
expertise for judicial authorities. It uses mainly biological tissues of the person as the dosimeter and
also can use materials from personal objects as well as those located in the immediate environment.
EPR dosimetry is based on the fundamental properties of ionizing radiation: the generation of unpaired
electron species (often but not exclusively free radicals) proportional to absorbed dose. The technique
of EPR specifically and sensitively detects the amount of unpaired electrons that have sufficient stability
to be observed after their generation; while the amount of the detectable unpaired electrons is usually
directly proportional to the amount that was generated, these species can react, and therefore, the
relationship between the intensity of the EPR signal and the radiation dose needs to be established for
each type of use. The most extensive use of the technique has been with calcified tissue, especially with
[15]
enamel from teeth. An IAEA technical report was published on the use for tooth enamel. To extend
the possibility of EPR retrospective dosimetry, new materials possibly suitable for EPR dosimetry are
regularly studied and associated protocols established. This International Standard is aimed to make
this technique more widely available, more easily applicable and useful for dosimetry. Specifically, this
International Standard proposes a methodological frame and recommendations to set up, validate, and
apply protocols from samples collection to dose reporting.
© ISO 2013 – All rights reserved v

---------------------- Page: 5 ----------------------
INTERNATIONAL STANDARD ISO 13304-1:2013(E)
Radiological protection — Minimum criteria for electron
paramagnetic resonance (EPR) spectroscopy for
retrospective dosimetry of ionizing radiation —
Part 1:
General principles
1 Scope
The primary purpose of this International Standard is to provide minimum acceptable criteria required
to establish procedure of retrospective dosimetry by electron paramagnetic resonance spectroscopy
and to report the results.
The second purpose is to facilitate the comparison of measurements related to absorbed dose estimation
obtained in different laboratories.
This International Standard covers the determination of absorbed dose in the measured material. It
does not cover the calculation of dose to organs or to the body. It covers measurements in both biological
and inanimate samples, and specifically:
a) based on inanimate environmental materials, usually made at X-band microwave frequencies
(8 GHz to 12 GHz);
b) in vitro tooth enamel using concentrated enamel in a sample tube, usually employing X-band
frequency, but higher frequencies are also being considered;
c) in vivo tooth dosimetry, currently using L-band (1 GHz to 2 GHz), but higher frequencies are also
being considered;
d) in vitro nail dosimetry using nail clippings measured principally at X-band, but higher frequencies
are also being considered;
e) in vivo nail dosimetry with the measurements made at X-band on the intact finger or toe;
f) in vitro measurements of bone, usually employing X-band frequency, but higher frequencies are also
being considered.
For the biological samples, the in vitro measurements are carried out in samples after their removal
from the person and under laboratory conditions, whereas the measurements in vivo may take place
under field conditions.
NOTE The dose referred to in this International Standard is the absorbed dose of ionizing radiation in the
measured materials.
2 Terms and definitions
For the purposes of this document, the following terms and definitions apply.
2.1
retrospective dosimetry (including early or emergency response)
dosimetry, usually at the level of the individual, carried out after an exposure using methods other than
the conventional radiation dosimeters
© ISO 2013 – All rights reserved 1

---------------------- Page: 6 ----------------------
ISO 13304-1:2013(E)

2.2
electron paramagnetic resonance (EPR)
electron spin resonance (ESR)
magnetic resonance technique which is similar to nuclear magnetic resonance (NMR) but based on the
net spin of unpaired electrons, such as free radicals and electron defects centers in matrices
Note 1 to entry: The terms EPR and ESR are essentially equivalent and are widely used. The term electron
magnetic resonance (EMR) also sometimes is used because it is analogous to nuclear magnetic resonance (NMR).
2.3
radical/paramagnetic centre
species with unpaired electron(s)
Note 1 to entry: Paired electrons have the same quantum state except for opposite spins; unpaired electrons
do not have a “partner” with the opposite spin. When the unpaired spin is on a molecule, it is usually termed a
radical; when the unpaired electron is in a matrix, it often is termed a paramagnetic centre.
2.4
in vivo measurement
measurement carried out within the living system, such as measurements of paramagnetic centres in
teeth within the mouth
2.5
in vitro measurement
measurement carried out on materials outside the organism
Note 1 to entry: The term ex vivo also has been used in the literature for sample measured in vitro but irradiated
within the organism.
2.6
quality assurance
planned and systematic actions necessary to provide adequate confidence that a process, measurement,
or service satisfies given requirements for quality
2.7
quality control
planned and systematic actions intended to verify that systems and components conform with
predetermined requirements
3 Confidentiality and ethical considerations
All individual identifying information of persons who provided samples should not be attached to the
information on the samples and kept only in a secured place. The corresponding samples should be
identified by codes with indication only of parameters that are useful for scientific purposes and for
making decisions. Data linking the code to the person can be kept if they are done so in a secure manner,
with access limited to the persons in charge of the data.
Where appropriate, permission for obtaining and measuring the samples should be obtained under the
rules of the jurisdiction where the samples are obtained.
4 Laboratory safety requirements
4.1 Magnetic field
With conventional spectrometers, the magnetic field (for signals with g-factor near 2,0, typically 350 mT
for X-band and 1 200 mT for Q-band) is restricted to the region between the poles of the magnets, and
therefore, there is no associated biological hazard (can affect watches or credit cards if brought very
close to the gap).
2 © ISO 2013 – All rights reserved

---------------------- Page: 7 ----------------------
ISO 13304-1:2013(E)

The open nature of some in vivo EPR spectrometers (for signals with g-factor near 2,0, 40 mT for L-band)
combined with large gaps between the poles has the potential to project the 0,5 mT line beyond the
confines of the room. This line needs to be determined and appropriate shielding placed for areas that
exceed this limit and that are accessed by the general public. The establishment of the 0,5 mT limit
is based on concerns about potential effects on pacemakers, which are the only significant source of
biohazards from the magnetic fields that are employed with EPR. The conventional limit is 0,5 mT
(which is very conservative) and surveys should be made to confirm that this field is not exceeded where
a person with a pacemaker could be positioned.
Effects of modulation fields on tissues or tooth restorations are not a significant hazard.
4.2 Electromagnetic frequency
4.2.1 in vitro measurement
The configurations used for in vitro measurements have no hazard for exposure of operators, as
the spectrometer usually fully constrains the microwave to the sample with no significant amount
distributed outside of the resonator.
4.2.2 in vivo measurement
Measurements in vivo have the potential hazard of local heating. The operative safety limit is that
established for NMR in terms of permissible rates of energy absorption. In practice, this is a potential
hazard only at high incident microwave power levels—typically > 1 W, which is at least a factor of 3
greater than that in existing instruments.
4.3 Biohazards from samples
Biological samples measured in vitro should be handled in conformance to the rules of the jurisdiction
for routine practice for handling biological samples.
Measurements of teeth in vivo should follow the routines practiced for ordinary dentistry in regard to
potential contamination from subjects to operators or other subjects.
5 Collection/selection and identification of samples
All samples should be collected in as uniform manner as possible and the circumstances of the collection
noted, although this may not always be able to be controlled by the measuring laboratory. If prior
coordination between the collecting and the measuring laboratories is possible, requirements about the
sample collection, selection (of donors, location, or materials) and storage (sample holder, integrity of
the sample and of the container, temperature, light, UV) should be given. If information about samples
is available, keep record of them (this information can be about the location of the sample, origin or
history of the sample, information about donor, etc.). All samples should have a unique identifying code
associated with them.
6 Transportation and storage of samples
If sample collection is made in a place other than the measuring laboratory, then samples should
be transported and stored under specified environmental conditions. These conditions should be
coordinated between the collecting and the measuring laboratories. Conditions of transportation and
storage of the sample may affect the integrity of the sample and also modify the quantity of paramagnetic
species or the nature of the paramagnetic species in the samples. Environmental parameters such
as light and other types of radiations (UV, X-rays, gamma), temperature, humidity, oxygen, sample
conditionings in water or disinfectant solution, for example, contamination (e.g. dust), may significantly
affect the nature and quantity of paramagnetic species in the samples. Therefore, specific attention
should be taken as to the conditions of transportation and storage to avoid or limit as much as possible
the influence of environmental parameters on the samples.
© ISO 2013 – All rights reserved 3

---------------------- Page: 8 ----------------------
ISO 13304-1:2013(E)

If possible, the influence of these parameters on the radiation-induced signal line shape and intensity
should be investigated to establish the optimum conditions for transportation or storage and to avoid
unnecessary precautions. When samples are known to be sensitive to one or several environmental
conditions or the influence of these parameters or samples is not known, it is highly recommended that
precautions are taken so as to avoid conditions that could affect the samples.
Transportation conditions, including dates, ways of transportation, and mode of control of transportation
conditions, should be recorded. Appropriate sample packaging should always be used to prevent sample
physical damage.
Procedures to avoid X-ray exposure of the sample during airport controls should be implemented. The
dose at the X-ray hand luggage control is of the order of the microgray, so it can be considered negligible
for some applications. If not, when the sample is transported in hand luggage, then authorization for
X-ray exemption should be obtained in advance in order to avoid hindrance at the airport security
controls. X-ray dose to the hold luggage can be higher. For shipping, appropriate labelling (including a
note that the package contains radiation-sensitive dosimeters and, therefore, should not be irradiated)
should be used. When this is not possible, unirradiated identical control samples or dosimeters should
be placed in the package.
After the samples are received, they should be stored under stable conditions and the temperature and
humidity should be monitored and recorded. Exposure to light should always be avoided.
7 Preparation of samples
Sample preparation should be performed according to an established and explicit protocol.
For in vitro and ex vivo measurements, sample preparation is usually needed to accomplish several goals,
including: achieving a sample size that fits in the measurement tube; reducing anisotropy; ensuring
disinfection; eliminating paramagnetic impurities from the sample; drying the sample; and stabilizing
the EPR signals.
When required, preparation of the sample can be done by grinding, crushing, cutting, drilling, or
other mechanical treatments. During these operations, sample overheating should be avoided using
water irrigation or other cooling systems. Metal contamination of the sample can be avoided by using
hard alloy tools.
As needed, sterilization, cleaning, deproteination, and/or delipidation are performed using chemical
agents. Thermal treatment (annealing, freezing) can be used to accelerate or slow down recombination
of the radicals. Samples with significant amounts of moisture can be dried before the EPR measurements
to improve signal-to-noise ratio.
The setup of a protocol for sample preparation shall include the evaluation of the effect of the protocol
on the EPR signals (lineshape and intensity) on the dose estimation, including whether it can induce EPR
signals. When employing the additive dose method (see 10.2.1), it is very desirable to use protocols that
do not affect the radiation sensitivity.
The protocol should be described in details in documents, including: the duration of treatment, quality of
reagents, and the instrumentation used and its performance. All samples should be prepared following
the same protocol. Samples used for calibration have to be treated according to the same protocol as the
samples to be measured.
Any modification to the protocol should be noted and the influence of each modification evaluated (e.g.
power or frequency of ultrasonic bath, reagent quality).
All details of the procedures for each sample shall be recorded in a log of the history of the sample.
For measurements in vivo, there are no requirements for preparation of the samples. Depending on the site
that is measured, there may be a need to minimize moisture (especially when making measurements in vivo
in teeth) or to carry out some cleaning procedures (e.g. removing obvious particulate matter from nails).
Because of the limited ability to control environmental conditions fully when making measurements in
4 © ISO 2013 – All rights reserved

---------------------- Page: 9 ----------------------
ISO 13304-1:2013(E)

vivo, it is highly desirable to always utilize a standard sample that is in place and with a known relationship
to the sample volume so that factors that affect the measurements (especially factors that affect the quality
factor of the resonator) can be detected and accounted for in the processing of the data.
8 Apparatus
8.1 Principles of EPR spectroscopy
EPR is a technique that specifically and sensitively detects unpaired electrons. It is based on the resonant
absorption of electromagnetic energy for transitions between electron spin states. A static magnetic
field is applied that induces net absorption from transitions between spin states if there is a vacant level
to which the spin can flip. In a magnetic field, the different spin states result in different energy levels,
with the difference in the energy being proportional to the magnetic field. A transition between these
two levels can be induced by an appropriate electromagnetic field.
Currently, continuous wave (CW) EPR spectroscopy is usually used for EPR dosimetry. In an EPR
CW spectrometer, the resonance frequency is applied to a resonant structure and absorption of the
electromagnetic waves by a sample in the resonator is detected. Typically, the resonant condition is
reached by continuously changing the main magnetic field, while a fixed frequency is applied to the
resonator. As a result, an EPR spectrum of absorption versus magnetic field intensity is obtained. Other
methods of EPR signal detection such as pulsed EPR, fast scan EPR spectroscopy, etc. are potentially
available, but to date, these have not been shown to be more effective for dosimetry application than
CW EPR. So, considerations on EPR dosimetry in this International Standard are restricted to CW EPR,
although most of the guidelines would be applicable to other types of EPR spectroscopy.
To improve the signal-to-noise ratio, modern EPR CW spectrometers employ high-frequency magnetic
field modulation in combination with phase-sensitive detection. As a result, the original spectral line
is produced not in the form of an absorption curve, but in the form of its first derivative. In modern
spectrometers, the EPR signal is recorded in digital form using a dedicated computer. In most
spectrometers, the computer also is used to control operation of the spectrometer, e.g. for setting
measurement parameters, tuning the resonator, acquiring the signal, saving the recorded spectrum to
disk, and preliminary spectra processing (such as digital filtering, baseline correction, etc.).
Depending on the magnetic field intensity and, respectively, the resonance frequency, the following
band frequencies are commonly used for EPR dosimetry.
— X-band usually is used for EPR in vitro dosimetry because of a good compromise between sensitivity,
sample size, and sensitivity to the presence of water.
— L-band is used mainly for in vivo tooth dosimetry because of the relatively low amount of non-
resonant absorption of the microwaves due to the presence of water in biological tissues. Q-band
is mainly used in research connected with investigation of spectroscopic properties of materials
suitable for EPR dosimetry and has potential for being utilized for in vitro dosimetry. An advantage
of Q-band is that only a small sample mass is required for measurements and spectral components
can be better resolved in comparison with lower frequencies. On the other hand, such spectrometers
are not widely available, often are more complex to use, and may have a lower signal-to-noise ratio.
8.2 Requirements for EPR spectrometers
As EPR dosimetry often deals with small sample masses and low intensity signals, the sensitivity and
stability of the instruments are critical. Sensitivity and stability may be optimized by proper choice of
instrumental factors (such as selection of resonator, its tuning, and minimization of the microphonic
effects) and selection of the measurement parameters.
8.3
...

NORME ISO
INTERNATIONALE 13304-1
Première édition
2013-07-01
Radioprotection — Critères minimaux
pour la spectroscopie par résonance
paramagnétique électronique (RPE)
pour la dosimétrie rétrospective des
rayonnements ionisants —
Partie 1:
Principes généraux
Radiological protection — Minimum criteria for electron
paramagnetic resonance (EPR) spectroscopy for retrospective
dosimetry of ionizing radiation —
Part 1: General principles
Numéro de référence
ISO 13304-1:2013(F)
©
ISO 2013

---------------------- Page: 1 ----------------------
ISO 13304-1:2013(F)

DOCUMENT PROTÉGÉ PAR COPYRIGHT
© ISO 2013
Droits de reproduction réservés. Sauf indication contraire, aucune partie de cette publication ne peut être reproduite ni utilisée
sous quelque forme que ce soit et par aucun procédé, électronique ou mécanique, y compris la photocopie, l’affichage sur
l’internet ou sur un Intranet, sans autorisation écrite préalable. Les demandes d’autorisation peuvent être adressées à l’ISO à
l’adresse ci-après ou au comité membre de l’ISO dans le pays du demandeur.
ISO copyright office
Case postale 56 • CH-1211 Geneva 20
Tel. + 41 22 749 01 11
Fax + 41 22 749 09 47
E-mail copyright@iso.org
Web www.iso.org
Publié en Suisse
ii © ISO 2013 – Tous droits réservés

---------------------- Page: 2 ----------------------
ISO 13304-1:2013(F)

Sommaire Page
Avant-propos .iv
Introduction .v
1 Domaine d’application . 1
2 Termes et définitions . 1
3 Confidentialité et considérations déontologiques . 2
4 Exigences de sécurité relatives aux laboratoires. 3
4.1 Champ magnétique . 3
4.2 Fréquence électromagnétique . 3
4.3 Risques biologiques pour les échantillons . 3
5 Prélèvement/choix et identification des échantillons . 3
6 Transport et stockage des échantillons . 4
7 Préparation des échantillons . 4
8 Appareillage . 5
8.1 Principes de la spectroscopie RPE . 5
8.2 Exigences relatives aux spectromètres RPE . 6
8.3 Exigences relatives au résonateur . 6
8.4 Mesurages des signaux non issus des échantillons . 6
8.5 Stabilité du spectromètre et surveillance/contrôle des conditions environnementales . 7
8.6 Dérive de la ligne de base . 7
9 Mesurages des échantillons . 8
9.1 Principes généraux . 8
9.2 Choix et optimisation des paramètres de mesure . 8
9.3 Positionnement et chargement de l’échantillon .10
9.4 Réglage du spectromètre.10
9.5 Utilisation d’échantillons de référence comme marqueurs de champ et
contrôleurs d’amplitude .11
9.6 Reproductibilité du contrôle .11
9.7 Procédure de mesure des échantillons anisotropes .11
9.8 Codage des spectres et des échantillons .11
10 Détermination de la dose absorbée dans les échantillons .11
10.1 Détermination de l’intensité du signal radio-induit .11
10.2 Conversion du signal RPE en une estimation de dose absorbée .12
11 Incertitude de mesure .13
12 Examen d’une dose suspecte.13
13 Assurance qualité et contrôle de la qualité (AQ et CQ) .14
14 Exigences minimales concernant la documentation .16
Bibliographie .17
© ISO 2013 – Tous droits réservés iii

---------------------- Page: 3 ----------------------
ISO 13304-1:2013(F)

Avant-propos
L’ISO (Organisation internationale de normalisation) est une fédération mondiale d’organismes
nationaux de normalisation (comités membres de l’ISO). L’élaboration des Normes internationales est
en général confiée aux comités techniques de l’ISO. Chaque comité membre intéressé par une étude
a le droit de faire partie du comité technique créé à cet effet. Les organisations internationales,
gouvernementales et non gouvernementales, en liaison avec l’ISO participent également aux travaux.
L’ISO collabore étroitement avec la Commission électrotechnique internationale (CEI) en ce qui concerne
la normalisation électrotechnique.
Les procédures utilisées pour élaborer le présent document et celles destinées à sa mise à jour sont
décrites dans les Directives ISO/CEI, Partie 1. Il convient, en particulier de prendre note des différents
critères d’approbation requis pour les différents types de documents ISO. Le présent document a été
rédigé conformément aux règles de rédaction données dans les Directives ISO/CEI, Partie 2, www.iso.
org/directives.
L’attention est appelée sur le fait que certains des éléments du présent document peuvent faire l’objet de
droits de propriété intellectuelle ou de droits analogues. L’ISO ne saurait être tenue pour responsable
de ne pas avoir identifié de tels droits de propriété et averti de leur existence. Les détails concernant les
références aux droits de propriété intellectuelle ou autres droits analogues identifiés lors de l’élaboration
du document sont indiqués dans l’Introduction et/ou sur la liste ISO des déclarations de brevets reçues,
www.iso.org/patents.
Les éventuelles appellations commerciales utilisées dans le présent document sont données pour
information à l’intention des utilisateurs et ne constituent pas une approbation ou une recommandation.
Le comité chargé de l’élaboration du présent document est l’ISO/TC 85, Énergie nucléaire, technologies
nucléaires et radioprotection, sous-comité SC 2, Radioprotection.
L’ISO 13304 comprend les parties suivantes, présentées sous le titre général Radioprotection — Critères
minimaux pour la spectroscopie par résonance paramagnétique électronique (RPE) pour la dosimétrie
rétrospective des rayonnements ionisants:
— Partie 1: Principes généraux
iv © ISO 2013 – Tous droits réservés

---------------------- Page: 4 ----------------------
ISO 13304-1:2013(F)

Introduction
La résonance paramagnétique électronique (RPE) est une technique couramment utilisée en dosimétrie
rétrospective lorsque les informations dosimétriques concernant un individu sont potentiellement
incomplètes ou inconnues. Elle est largement utilisée pour l’évaluation rétrospective des doses délivrées
à des moments précis dans le passé (par exemple, la dosimétrie RPE est l’une des méthodes de choix pour
l’évaluation rétrospective des doses délivrées aux populations exposées à l’arme atomique au Japon et
aux personnes exposées suite à l’accident de Tchernobyl) et elle est également envisagée comme méthode
de tri de population lors d’accident impliquant un grand nombre de personnes potentiellement exposées
à des niveaux de rayonnement cliniquement significatifs. Divers types de matériaux peuvent être
analysés par la technique RPE pour estimer la dose absorbée dans ces matériaux. Ainsi, la spectroscopie
RPE est un outil polyvalent pour la dosimétrie rétrospective, qui est aussi bien mis en œuvre pour les
expositions aiguës (passées ou récentes, d’une partie ou de l’ensemble du corps) que pour les expositions
de longue durée. Les doses estimées avec la méthode RPE ont principalement ont été utilisées pour
établir une corrélation entre les effets biologiques des rayonnements ionisants et la dose reçue, pour
valider d’autres techniques ou méthodologies, pour la gestion des victimes d’accident d’irradiation ou
assurer l’expertise médico-légale dans le cadre de procédures judiciaires. Les tissus biologiques humains
sont les principaux matériaux utilisés pour la dosimétrie rétrospective par RPE mais des matériaux
provenant d’objets personnels ainsi que des objets situés dans l’environnement immédiat des personnes
exposées peuvent être également utilisés. Le principe de la dosimétrie RPE est basée sur les propriétés
fondamentales des rayonnements ionisants: la production d’espèces comportant des électrons non
appariés (souvent, mais non exclusivement, des radicaux libres) en quantité proportionnelle à la dose
absorbée. La spectroscopie RPE permet de détecter de manière spécifique et sensible les quantités
d’espèces comportant des électrons non appariés suffisamment stables pour pouvoir être observés
après leur création; bien que la quantité d’électrons non appariés détectés soit en général directement
proportionnelle à la quantité générée, ces espèces peuvent réagir, d’où la nécessité d’établir, pour chaque
type d’utilisation, la relation entre l’intensité du signal RPE et la dose de rayonnement. Cette technique a
été le plus souvent appliquée sur des tissus calcifiés, notamment l’émail dentaire. Un Rapport technique
de l’AIEA portant sur l’estimation des doses à partir de la mesure de l’émail dentaire a été publié en
[15]
2002. Pour étendre le champ d’application de la spectroscopie RPE dans le domaine de la dosimétrie
rétrospective, de nouveaux matériaux potentiellement utilisables sont régulièrement étudiés et des
protocoles associés établis. Le but de la présente norme est de faciliter la diffusion de cette technique,
de la rendre plus facilement applicable et plus utile pour la dosimétrie. Spécifiquement, la présente
Norme internationale propose un cadre méthodologique et des recommandations pour établir, valider
et appliquer des protocoles allant du prélèvement des échantillons jusqu’à la consignation des doses.
© ISO 2013 – Tous droits réservés v

---------------------- Page: 5 ----------------------
NORME INTERNATIONALE ISO 13304-1:2013(F)
Radioprotection — Critères minimaux pour la spectroscopie
par résonance paramagnétique électronique (RPE) pour la
dosimétrie rétrospective des rayonnements ionisants —
Partie 1:
Principes généraux
1 Domaine d’application
Le but principal de la présente Norme internationale est de fournir un ensemble de critères minimaux
acceptables requis pour établir une procédure de dosimétrie rétrospective par spectroscopie par
résonance paramagnétique électronique et pour rendre compte des résultats.
Le but secondaire de la présente Norme internationale vise à faciliter la comparaison entre laboratoire
des mesurages et de l’estimation de la dose absorbée.
La présente Norme internationale couvre la détermination de la dose absorbée dans le matériau mesuré.
Elle ne couvre pas le calcul de la dose délivrée aux organes ou à l’organisme entier. Elle ne concerne
que les mesurages effectuées sur des échantillons biologiques et des échantillons inertes, et plus
particulièrement:
a) les mesurages de matériaux environnementaux inertes, généralement réalisés des fréquences
micro-ondes de la bande X (8 GHz à 12 GHz);
b) les mesurages in vitro de prélèvement d’émail dentaire, placé dans un tube porte-échantillon, et
mesuré en général en bande X, mais l’utilisation de fréquences micro-ondes plus élevées peut être
également considérée;
c) les mesurages in vivo de dents, réalisés actuellement en bande L (1 GHz à 2 GHz), mais des fréquences
micro-ondes plus élevées sont également envisagées;
d) les mesurages in vitro de prélèvements d’ongles effectués principalement dans la bande X, mais des
fréquences micro-ondes plus élevées sont également à l’étude;
e) les mesurages in vivo des ongles, effectués en bande X sur les ongles des doigts ou des orteils;
f) les mesurages in vitro de tissus osseux, réalisés en général en bande X mais l’utilisation de fréquences
micro-ondes plus élevées est également étudiée.
En ce qui concerne les échantillons biologiques, les mesurages in vitro sont effectués sur des échantillons
prélevés sur la personne et dans des conditions de laboratoire, tandis que les systèmes de mesure in
vivo, réalisés sur les individus peuvent être déplacés au plus près des victimes.
NOTE La dose mentionnée dans la présente Norme internationale, est la dose absorbée de rayonnement
ionisant dans les matériaux mesurés.
2 Termes et définitions
Pour les besoins du présent document, les termes et définitions suivants s’appliquent.
© ISO 2013 – Tous droits réservés 1

---------------------- Page: 6 ----------------------
ISO 13304-1:2013(F)

2.1
dosimétrie rétrospective (inclus les situations d’urgence)
dosimétrie, effectuée après une exposition aux rayonnements ionisants en utilisant des méthodes et
des matériaux autres que les méthodes classiques de surveillance dosimétrique pour estimer une dose
reçue par un individu
2.2
RPE (résonance paramagnétique électronique)
RSE (résonance de spin électronique)
technique de résonance magnétique qui est similaire à la résonance magnétique nucléaire (RMN), mais
fondée sur la résonance des spin électroniques non appariés, tel que par exemple ceux présents dans les
radicaux libres ou ceux liés à une vacance ou une substitution d’atomes dans un réseau cristallin
Note 1 à l’article: Le terme de RPE est maintenant le plus communément utilisé. Le terme RSE est de moins en
moins employé. Le terme résonance magnétique électronique (RME), plus récent, est parfois employé car il est
analogue au terme RMN (résonance magnétique nucléaire).
2.3
centre paramagnétique
espèce contenant un (des) électron(s) non apparié(s)
Note 1 à l’article: Les électrons appariés ont le même état quantique, mais des spins orientés de manière opposée;
les électrons non appariés n’ont pas de «partenaire» avec un spin électronique opposé. Lorsque le spin non apparié
se trouve sur une molécule, il est habituellement désigné par «radical»; lorsque l’électron non apparié se trouve
dans une matrice, il est souvent désigné par «défaut paramagnétique».
2.4
mesurage in vivo
mesurage réalisé sur un système vivant comme, par exemple, des mesurages de centres paramagnétiques
présents dans l’émail dentaire mesurés directement dans la cavité buccale
2.5
mesurage in vitro
mesurage de prélèvements biologiques réalisé à l’extérieur de l’organisme
Note 1 à l’article: Le terme ex vivo a également été utilisé dans la littérature pour des échantillons prélevés et
analysés in vitro, mais irradiés à l’intérieur de l’organisme.
2.6
assurance qualité
toutes les actions planifiées et systématiques nécessaires pour attester qu’un processus, un mesurage
ou un service satisfait aux exigences de qualité
2.7
contrôle de la qualité
actions planifiées et systématiques destinées à vérifier que les systèmes et les composants sont
conformes aux exigences prédéterminées
3 Confidentialité et considérations déontologiques
Aucune information permettant d’identifier un donneur d’échantillons ne doit être jointe aux
informations figurant sur les échantillons. Ces informations doivent être sécurisées. Il convient
d’identifier les échantillons correspondants par des codes avec seulement une indication des paramètres
ayant un intérêt scientifique et être utile pour mener l’analyse. La conservation des données reliant le
code à l’identité de la personne est autorisée à condition qu’elle soit sécurisée, avec un accès limité aux
personnes en charge des données.
Le cas échéant, il convient que la permission pour l’acquisition et le mesurage des échantillons soit
obtenue conformément à la législation du lieu d’obtention des échantillons.
2 © ISO 2013 – Tous droits réservés

---------------------- Page: 7 ----------------------
ISO 13304-1:2013(F)

4 Exigences de sécurité relatives aux laboratoires
4.1 Champ magnétique
Avec les spectromètres conventionnels, le champ magnétique (pour les signaux avec un facteur g proche
de 2,0, typiquement 350 mT en bande X, et 1 200 mT en bande Q) est confiné dans l’espace entre les
pôles des aimants, l’entrefer, il n’y a donc a priori aucun risque biologique associé lors de l’utilisation
(Toutefois, les montres ou les cartes à lecture magnétique sont endommagées lorsque placées dans et à
proximité de l’entrefer).
L’absence de confinement du champ magnétique de certains spectromètres RPE in vivo (pour les
signaux avec un facteur g proche de 2,0, 40 mT en bande L) associée aux espacements importants entre
les pôles peut potentiellement induire des champs supérieurs à la limite admise de 0,5 mT au-delà du
périmètre du local du spectromètre. Il convient de s’assurer que cette limite n’est pas dépassée pour
les zones accessibles au grand public et en cas de dépassement un blindage approprié doit être installé.
L’établissement de la limite à 0,5 mT est fondé sur les effets potentiels sur les pacemakers qui sont
l’unique source significative de risques engendrés par les champs magnétiques utilisés avec la RPE. La
limite conventionnelle est fixée à 0,5 mT (qui est une valeur très conservative) et il convient que des
contrôles soient effectués pour confirmer que cette limite n’est pas dépassée si une personne équipée
d’un pacemaker est susceptible de se trouver à proximité.
Pour les mesurages in vivo, les effets des champs magnétiques sur les tissus ou des restaurations
dentaires ne constituent pas un risque significatif.
4.2 Fréquence électromagnétique
4.2.1 Mesurage in vitro
Les configurations utilisées pour les mesurages in vitro ne présentent pas de risque connu concernant
l’exposition des opérateurs puisque les micro-ondes sont concentrées sur l’échantillon sans qu’une
quantité notable ne soit émise à l’extérieur du résonateur.
4.2.2 Mesurage in vivo
Les mesurages in vivo présentent le risque potentiel d’un échauffement local. La limite de sécurité fonctionnelle
est celle établie pour la RMN concernant les taux admissibles d’absorption d’énergie. Dans la pratique, il n’y a
de risque potentiel qu’à des niveaux élevés de puissance hyperfréquence incidente — habituellement > 1 W,
ce qui représente au moins un facteur trois fois supérieur à celui des instruments existants.
4.3 Risques biologiques pour les échantillons
Il convient que les échantillons d’origine biologique mesurés in vitro soient traités conformément aux
règles normalement applicables pour la manipulation des échantillons biologiques.
Il convient que les mesurages effectués sur des dents in vivo soient conformes aux règles normalement
applicables chez les professionnels en dentaire eu égard à la contamination potentielle des opérateurs
ou autres personnes par les sujets examinés.
5 Prélèvement/choix et identification des échantillons
Il convient que les échantillons soient prélevés d’une manière aussi uniforme que possible et que les
conditions de prélèvement soient consignées, bien que le laboratoire de mesure ne soit pas toujours en
mesure de le contrôler. Si une coordination préalable est possible entre les laboratoires en charge des
prélèvements et ceux réalisant les mesurages, il convient que les exigences concernant les prélèvements,
les règles de sélection (des donneurs, du lieu ou des matériaux) et la conservation (porte-échantillon,
intégrité de l’échantillon et du récipient, température, lumière, rayonnement ultraviolet) soient
spécifiées. Si des informations concernant les échantillons sont disponibles, il convient de les conserver
© ISO 2013 – Tous droits réservés 3

---------------------- Page: 8 ----------------------
ISO 13304-1:2013(F)

(il peut s’agir d’informations concernant l’emplacement, l’origine ou l’historique de l’échantillon, ou
d’informations concernant le donneur, etc.). Il convient qu’un code d’identification unique soit associé à
chacun des échantillons.
6 Transport et stockage des échantillons
Si le prélèvement des échantillons a lieu dans un endroit autre que le laboratoire de mesure, alors, il
convient que ces échantillons soient transportés et conservés dans des conditions environnementales
spécifiées. Il convient que ces conditions soient convenues entre les laboratoires en charge du prélèvement
et ceux en charge du mesurage. Les conditions de transport et de conservation de l’échantillon peuvent
altérer l’intégrité de l’échantillon et modifier la quantité ou la nature des espèces paramagnétiques dans
les échantillons. Les paramètres environnementaux, tels que la lumière et autres types de rayonnements
(UV, rayons X, gamma, etc.), la température, l’humidité, la teneur en oxygène, le conditionnement des
échantillons dans l’eau ou dans une solution désinfectante par exemple, ou encore la contamination
(par exemple poussière), peuvent altérer considérablement la nature et la quantité des espèces
paramagnétiques dans les échantillons. En conséquence, il convient de prêter une attention particulière
aux conditions de transport et de conservation afin d’éviter ou de limiter autant que possible l’influence
des paramètres environnementaux sur les échantillons.
Si possible, il convient d’étudier l’influence de ces paramètres sur la forme et l’intensité de la raie du signal
induit par les rayonnements ionisants afin de déterminer les conditions optimales pour le transport ou
la conservation et d’éviter les précautions inutiles. Si les échantillons sont réputés sensibles à une ou
plusieurs conditions environnementales, ou si l’influence de ces paramètres sur les échantillons n’est
pas connue, il est vivement recommandé que des précautions soient prises pour éviter les conditions
susceptibles d’affecter les échantillons.
Il convient que les conditions de transport, y compris les dates, les moyens de transport et le mode de
contrôle des conditions de transport soient consignées. Il convient que des emballages pour échantillons
soient toujours utilisés afin de protéger les échantillons contre toute détérioration physique.
Il convient que des procédures soient mises en œuvre pour éviter l’exposition des échantillons aux
rayons X lors des contrôles aux aéroports. La dose, au niveau du contrôle par rayons X des bagages à main,
étant de l’ordre du microgray, elle peut être considérée comme négligeable pour certaines applications.
Sinon, lorsque l’échantillon est transporté dans un bagage à main, il convient d’obtenir par avance une
autorisation dispensant le bagage du contrôle par rayons X afin d’éviter les soucis lors des contrôles de
sécurité dans les aéroports. Le bagage porté peut être soumis à une dose plus élevée de rayons X. Pour
l’expédition, il convient d’utiliser un étiquetage approprié (comprenant une note indiquant que le colis
contient des dosimètres sensibles aux rayonnements et qu’il convient de ne pas l’irradier). Si cela n’est
pas possible, il convient de placer des échantillons ou des dosimètres témoins identiques non irradiés
dans le colis.
Une fois les échantillons réceptionnés, il convient de les conserver dans des conditions stables et de
surveiller et de consigner la température et l’humidité. Il convient de toujours éviter l’exposition des
échantillons à la lumière.
7 Préparation des échantillons
Il convient que la préparation des échantillons soit effectuée selon un protocole établi et explicite.
Pour les mesurages in vitro et ex vivo, la préparation des échantillons est habituellement requise pour
atteindre plusieurs objectifs, parmi lesquels: l’obtention d’une taille d’échantillon adaptée au tube de
mesure, la réduction de l’anisotropie, la désinfection, l’élimination des impuretés paramagnétiques
contenues dans l’échantillon, le séchage de l’échantillon et la stabilisation des signaux RPE.
Si nécessaire, la préparation de l’échantillon peut être effectuée par broyage, concassage, découpage,
perçage ou autres traitements mécaniques. Durant ces opérations, il convient d’éviter de surchauffer
l’échantillon en utilisant un dispositif d’aspersion d’eau ou d’autres systèmes de refroidissement. La
contamination de l’échantillon par des métaux peut être évitée en utilisant des outils en alliages durs.
4 © ISO 2013 – Tous droits réservés

---------------------- Page: 9 ----------------------
ISO 13304-1:2013(F)

Selon les besoins, la stérilisation, le nettoyage, la déprotéinisation et/ou le dégraissage sont effectués
à l’aide d’agents chimiques. Il est possible de procéder à un traitement thermique (recuit, congélation)
pour accélérer ou ralentir la recombinaison des radicaux. Les échantillons présentant des taux
d’humidité élevés peuvent être séchés avant d’être soumis aux mesurages RPE afin d’améliorer le
rapport signal/bruit.
La mise en place d’un protocole de préparation des échantillons doit inclure l’évaluation de l’effet du
protocole sur les signaux RPE (forme de raie et intensité) et sur l’estimation de la dose, y compris
un possible effet induit sur les signaux RPE. Lors de l’utilisation de laméthode des doses additives
(voir 10.2.1), il est vivement recommandé d’utiliser des protocoles qui n’affectent pas la sensibilité aux
rayonnements.
Il convient que le protocole soit décrit en détail dans les documents et qu’il indique les éléments suivants:
la durée de traitement, la qualité des réactifs ainsi que l’instrumentation utilisée et ses performances.
Il convient que tous les échantillons soient préparés selon le même protocole. Les échantillons utilisés
pour l’étalonnage doivent être traités selon le même protocole que les échantillons à mesurer.
Il convient que toute modification du protocole soit consignée et que l’influence de chaque modification
soit évaluée (par exemple puissance ou fréquence du bain à ultrasons, qualité des réactifs).
Tous les détails des procédures relatives à chaque échantillon doivent être enregistrés dans un registre
répertoriant l’historique de l’échantillon.
Pour les mesurages in vivo, il n’existe aucune exigence concernant la préparation des échantillons. Selon
le site mesuré, il est possible qu’il soit nécessaire de réduire le taux d’humidité (notamment lors de
mesurages effectués in vivo sur des dents) ou p
...