Medical electrical equipment — Part 2-12: Particular requirements for basic safety and essential performance of critical care ventilators

ISO 80601-2-12:2011 applies to the basic safety and essential performance of a ventilator in combination with its accessories, hereafter referred to as me equipment: intended to be attended by a professional operator for those patients who are dependent on mechanical ventilation; and intended for use in critical care environments in a professional healthcare facility or intended for use in transport within a professional healthcare facility. ISO 80601-2-12:2011 is also applicable to those accessories intended by their manufacturer to be connected to a breathing system, or to a ventilator, where the characteristics of those accessories can affect the basic safety or essential performance of the ventilator. ISO 80601-2-12:2011 is not applicable to me equipment or an me system operating in ventilation modes intended for patients who are not dependent on mechanical ventilation. ISO 80601-2-12:2011 is not applicable to me equipment or an me system intended for those patients who are not dependent on mechanical ventilation. If a clause or subclause is specifically intended to be applicable to me equipment only, or to me systems only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to me equipment and to me systems, as relevant. ISO 80601-2-12:2011 is not applicable to continuous positive airway pressure (CPAP) me equipment, sleep apnoea therapy me equipment, home healthcare environment ventilators, ventilatory support me equipment, emergency and transport ventilators, anaesthetic ventilators, high-frequency jet ventilators (HFJVs) and high-frequency oscillatory ventilators (HFOVs). ISO 80601-2-12:2011 does not specify the requirements for me equipment that is intended solely to augment the ventilation of spontaneously breathing patients within a professional healthcare facility. ISO 80601-2-12:2011 does not specify the requirements for ventilators or accessories intended for anaesthetic applications which are given in ISO 80601-2-13. ISO 80601-2-12:2011 does not specify the requirements for ventilators or accessories intended for home care ventilators for ventilator-dependent patients which are given in ISO 10651-2. ISO 80601-2-12:2011 does not specify the requirements for ventilators or accessories intended for emergency and transport which are given in ISO 10651-3. ISO 80601-2-12:2011 does not specify the requirements for ventilators or accessories intended for home-care ventilatory support devices which are given in ISO 10651-6.

Appareils électromédicaux — Partie 2-12: Exigences particulières relatives à la sécurité de base et aux performances essentielles des ventilateurs pulmonaires pour utilisation en soins intensifs

L'ISO 80601-2-12:2010 s'applique à la sécurité de base et aux performances essentielles d'un ventilateur associé à ses accessoires, ci-après désignés par appareil em: destinés à être utilisés par un opérateur professionnel sur des patients dépendants d'une ventilation mécanique; et destinés à être utilisés dans un environnement de soins intensifs, dans un établissement de santé professionnel ou lors d'un transport dans un établissement de soins professionnel. L'ISO 80601-2-12:2010 s'applique également aux accessoires conçus par leur fabricant pour être connectés à un système respiratoire ou un ventilateur, lorsque les caractéristiques de ces accessoires peuvent affecter la sécurité de base et les performances essentielles du ventilateur. L'ISO 80601-2-12:2010 n'est applicable à aucun appareil em ni système em fournissant des modes de ventilation pour patients non dépendants de la ventilation mécanique. L'ISO 80601-2-12:2010 n'est applicable à aucun appareil em ni système em pour patient non dépendant de la ventilation mécanique. Si un article ou un paragraphe est spécifiquement destiné à être applicable uniquement aux appareils em ou uniquement aux systèmes em, le titre et le contenu de cet article ou de ce paragraphe l'indiquent. Si cela n'est pas le cas, l'article ou le paragraphe s'applique aux appareils em et aux systèmes em, selon le cas. L'ISO 80601-2-12:2010 ne s'applique pas aux appareils em délivrant une pression positive continue (CPAP), aux appareils em de thérapie d'apnée du sommeil, aux ventilateurs de soins de santé à domicile, aux appareils em d'assistance respiratoire, aux ventilateurs de secours et de transport, aux ventilateurs d'anesthésie, aux jets ventilateurs à haute-fréquence (JVHF) et aux ventilateurs à oscillation haute-fréquence (VOHF). L'ISO 80601‑2‑12:2010 ne précise pas les exigences liées aux appareils em uniquement destinés à augmenter la ventilation des patients respirant spontanément au sein d'un établissement de santé professionnel. L'ISO 80601-2-12:2010 ne précise pas les exigences liées aux ventilateurs ou accessoires prévus pour des applications anesthésiques. Celles-ci sont données dans l'ISO 80601-2-13. L'ISO 80601-2-12:2010 ne précise pas les exigences liées aux ventilateurs ou accessoires destinés aux ventilateurs pour soins à domicile pour patients dépendants d'un ventilateur. Celles-ci sont données dans l'ISO 10651-2. L'ISO 80601-2-12:2010 ne précise pas les exigences liées aux ventilateurs ou accessoires prévus pour les urgences et le transport. Celles-ci sont données dans l'ISO 10651-3. L'ISO 80601-2-12:2010 ne précise pas les exigences liées aux ventilateurs ou accessoires prévus pour les dispositifs d'assistance respiratoire à domicile. Celles-ci sont données dans l'ISO 10651-6.

General Information

Status
Withdrawn
Publication Date
06-Apr-2011
Withdrawal Date
06-Apr-2011
Current Stage
9599 - Withdrawal of International Standard
Completion Date
25-Feb-2020
Ref Project

Relations

Buy Standard

Standard
ISO 80601-2-12:2011 - Medical electrical equipment
English language
78 pages
sale 15% off
Preview
sale 15% off
Preview
Standard
ISO 80601-2-12:2011 - Appareils électromédicaux
French language
82 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)

INTERNATIONAL ISO
STANDARD 80601-2-12
First edition
2011-04-15


Medical electrical equipment —
Part 2-12:
Particular requirements for basic safety
and essential performance of critical care
ventilators
Appareils électromédicaux —
Partie 2-12: Exigences particulières relatives à la sécurité de base et
aux performances essentielles des ventilateurs pulmonaires pour
utilisation en soins intensifs




Reference number
ISO 80601-2-12:2011(E)
©
ISO 2011

---------------------- Page: 1 ----------------------
ISO 80601-2-12:2011(E)

COPYRIGHT PROTECTED DOCUMENT


©  ISO 2011
All rights reserved. Unless otherwise specified, no part of this publication may be reproduced or utilized in any form or by any means,
electronic or mechanical, including photocopying and microfilm, without permission in writing from either ISO at the address below or
ISO's member body in the country of the requester.
ISO copyright office
Case postale 56  CH-1211 Geneva 20
Tel. + 41 22 749 01 11
Fax + 41 22 749 09 47
E-mail copyright@iso.org
Web www.iso.org
Published in Switzerland

ii © ISO 2011 – All rights reserved

---------------------- Page: 2 ----------------------
ISO 80601-2-12:2011(E)
Contents Page
Foreword . vi
Introduction . viii
201.1 Scope, object and related standards . 1
201.1. 1 Scope . 1
201.1. 2 Object . 2
201.1. 3 Collateral standards . 2
201.1. 4 Particular standards . 2
201.2 Normative references . 3
201.3 Terms and definitions . 6
201.4 General requirements . 9
201.4. 3 ESSENTIAL PERFORMANCE . 9
201.4. 3.101 * Additional requirements for ESSENTIAL PERFORMANCE . 9
201.4. 6 * ME EQUIPMENT or ME SYSTEM parts that contact the PATIENT . 9
201.4. 11.101 * Additional requirements for pressurized gas input . 9
201.4. 11.101.1 Overpressure requirement . 9
201.4. 11.101.2 Compatibility requirement . 10
201.5 General requirements for testing of ME EQUIPMENT . 10
201.5. 101 * Additional requirements for general requirements for testing of ME EQUIPMENT . 10
201.5. 101.1 VENTILATOR test conditions . 10
201.5. 101.2 * Gas flowrate and leakage specifications . 11
201.5. 101.3 * VENTILATOR testing errors . 11
201.6 Classification of ME EQUIPMENT and ME SYSTEMS . 11
201.7 ME EQUIPMENT identification, marking and documents . 11
201.7. 2.3 * Consult ACCOMPANYING DOCUMENTS . 11
201.7. 2.101 Additional requirements for marking on the outside of ME EQUIPMENT or ME EQUIPMENT
parts . 11
201.7. 2.4.101 Additional requirements for ACCESSORIES . 12
201.7. 2.13.101 Additional requirements for physiological effects . 12
201.7. 2.17.101 Additional requirements for protective packaging . 13
201.7. 4.3 * Unit of measure . 13
201.7. 9.1 Additional general requirements . 13
201.7. 9.2.1.101 Additional general requirements . 14
201.7. 9.2.2.101 * Additional requirements for warnings and safety notices . 14
201.7. 9.2.8.101 * Additional requirements for start-up procedure. 15
201.7. 9.2.9.101 * Additional requirements for operating instructions . 15
201.7. 9.2.12 Cleaning, disinfection, and sterilization . 16
201.7. 9.2.14.101 * Additional requirements for ACCESSORIES, supplementary equipment, used
material . 16
201.7. 9.2.16.101 * Additional requirements for reference to the technical description . 16
201.7. 9.3.1.101 * Additional general requirements . 16
201.7. 9.3.101 Additional requirements for the technical description . 17
201.8 Protection against electrical HAZARDS from ME EQUIPMENT . 17
201.9 Protection against mechanical HAZARDS of ME EQUIPMENT and ME SYSTEMS . 17
201.9. 6.2.1.101 Additional requirements for audible acoustic energy . 17
201.9. 101 * Additional requirements for suction procedures . 18
201.10 Protection against unwanted and excessive radiation HAZARDS . 20
© ISO 2011 – All rights reserved iii

---------------------- Page: 3 ----------------------
ISO 80601-2-12:2011(E)
201.11 Protection against excessive temperatures and other HAZARDS .20
201.11. 6.4 Leakage .20
201.11. 6.5.101 * Additional requirements for ingress of water or particulate matter
into ME EQUIPMENT or ME SYSTEM .21
201.11. 6.6 * Cleaning and disinfection of ME EQUIPMENT or ME SYSTEM .21
201.11. 6.7 Sterilization of ME EQUIPMENT or ME SYSTEM .21
201.11. 8.101 Additional requirements for interruption of the power supply/SUPPLY MAINS
to ME EQUIPMENT .22
201.11. 8.101.1 TECHNICAL ALARM CONDITION for power supply failure .22
201.11. 8.101.2 INTERNAL ELECTRICAL POWER SOURCE or external reserve electrical power source .22
201.12 Accuracy of controls and instruments and protection against hazardous outputs .23
201.12. 1 * Accuracy of controls and instruments .23
201.12. 1.101 Volume-controlled breath type .23
201.12. 1.102 Pressure-controlled breath type .26
201.12. 1.103 * DELIVERED VOLUME MONITORING .29
201.12. 1.104 * Response of the VENTILATOR to an increase in O concentration .29
2
201.12. 4 Protection against hazardous output .31
201.12. 4.101 Oxygen monitor .31
201.12. 4.102 * Measurement of AIRWAY PRESSURE .31
201.12. 4.103 * Measurement of expired volume and low-volume ALARM CONDITIONS .32
201.12. 4.103.1 VENTILATORS intended to provide a DELIVERED VOLUME > 50 ml .32
201.12. 4.103.2 VENTILATORS intended to provide a DELIVERED VOLUME u 50 ml .32
201.12. 4.104 * MAXIMUM LIMITED PRESSURE PROTECTION DEVICE .33
201.12. 4.105 High-pressure ALARM CONDITION and PROTECTION DEVICE .33
201.12. 4.106 PEEP ALARM CONDITIONS.33
201.12. 4.107 * Obstruction ALARM CONDITION.33
201.12. 4.108 * Partial-occlusion ALARM CONDITION .34
201.12. 101 * Protection against accidental adjustments .34
201.13 HAZARDOUS SITUATIONS and fault conditions .34
201.13. 2.101 * Additional specific SINGLE FAULT CONDITIONS .34
201.13. 102 * Failure of one gas supply to a VENTILATOR .35
201.13. 103 * Independence of ventilation control function and related RISK CONTROL measures .35
201.14 PROGRAMMABLE ELECTRICAL MEDICAL SYSTEMS (PEMS) .35
201.14. 101 Software life cycle .35
201.15 Construction of ME EQUIPMENT .35
201.15. 3.5.101 Additional requirements for rough handling .35
201.15. 3.5.101.1 * Shock and vibration .35
201.15. 3.5.101.2 * Shock and vibration for a MOBILE VENTILATOR .36
201.15. 101 Mode of operation .37
201.15. 102 Delivered Oxygen concentration .38
201.15. 103 ACCESSORY self-check .38
201.16 ME SYSTEMS .38
201.16. 1.101 Additional general requirements for ME SYSTEMS .38
201.17 Electromagnetic compatibility of ME EQUIPMENT and ME SYSTEMS .38
201.17. 101 Additional requirements for electromagnetic compatibility of ME EQUIPMENT and
ME SYSTEMS .38
201.101 Gas connections .38
201.101. 1 * Protection against reverse gas leakage .38
201.101. 2 Connection to the MEDICAL GAS PIPELINE SYSTEM .39
201.101. 3 VBS connectors .39
201.101. 3.1 * General .39
201.101. 3.2 Other named ports .39
201.101. 3.2.1 PATIENT-CONNECTION PORT .39
201.101. 3.2.2 GAS OUTPUT PORT and GAS RETURN PORT .39
201.101. 3.2.3 MANUAL ventilation port .39
201.101. 3.2.4 EMERGENCY INTAKE PORT .40
iv © ISO 2011 – All rights reserved

---------------------- Page: 4 ----------------------
ISO 80601-2-12:2011(E)
201.101. 3.2.5 FLOW-DIRECTION-sensitive components . 40
201.101. 3.2.6 ACCESSORY port . 40
201.101. 3.2.7 Monitoring probe port . 40
201.101. 3.2.8 Gas EXHAUST PORT . 40
201.102 Requirements for the VBS and ACCESSORIES . 40
201.102. 1 * General . 40
201.102. 2 Labelling . 40
201.102. 3 Breathing tubes . 41
201.102. 4 * Water management . 41
201.102. 4.1 Humidification system . 41
201.102. 4.2 HEAT AND MOISTURE EXCHANGER (HME) . 41
201.102. 5 Gas mixers . 41
201.102. 6 BREATHING SYSTEM FILTERS . 41
201.102. 7 VENTILATOR BREATHING SYSTEMS . 41
201.102. 7.1 Leakage from complete VBS . 41
201.102. 7.2 * Non-invasive ventilation . 42
201.103 * Spontaneous breathing during loss of power supply . 42
201.104 * Training . 43
201.105 * Indication of duration of operation . 43
201.106 SIGNAL INPUT/OUTPUT PART . 43
201.106. 1 General . 43
201.106. 2 * Connection to an electronic health record . 43
201.106. 3 * Connection to a DISTRIBUTED ALARM SYSTEM . 43
201.106. 4 * Connection for remote control . 43
201.107 Display loops . 43
201.107. 1 Pressure-volume loops . 43
201.107. 2 Flow-volume loops . 44
201.108 * Timed ventilatory pause . 44
201.108. 1 Expiratory pause . 44
201.108. 2 Inspiratory pause. 44
202 Medical electrical equipment – Part 1-2: General requirements for basic safety and
essential performance – Collateral standard: Electromagnetic compatibility –
Requirements and tests . 45
202.6.2.1.10 * Compliance criteria . 45
206 Medical electrical equipment – Part 1-6: General requirements for basic safety and
essential performance – Collateral Standard: Usability . 46
208 Medical electrical equipment – Part 1-8: General requirements for basic safety and
essential performance – Collateral Standard: General requirements, tests and guidance
for alarm systems in medical electrical equipment and medical electrical systems . 47
208.6.3.3.2.101 * Additional requirements for characteristics of ALARM CONDITION logging . 47
208.6.8.3.101 Additional requirements for global indefinite ALARM SIGNAL inactivation states . 47
208.6.8.4.101 * Additional requirements for termination of ALARM SIGNAL inactivation . 47
Annex C (informative) Guide to marking and labelling requirements for ME EQUIPMENT and
ME SYSTEMS . 48
Annex D (informative) Symbols on marking . 53
Annex AA (informative) Particular guidance and rationale . 55
Annex BB (informative) Reference to the Essential Principles . 72
Bibliography . 74
Alphabetized index of defined terms used in this particular standard . 77

© ISO 2011 – All rights reserved v

---------------------- Page: 5 ----------------------
ISO 80601-2-12:2011(E)
Foreword
ISO (the International Organization for Standardization) is a worldwide federation of national standards bodies
(ISO member bodies). The work of preparing International Standards is normally carried out through ISO
technical committees. Each member body interested in a subject for which a technical committee has been
established has the right to be represented on that committee. International organizations, governmental and
non-governmental, in liaison with ISO, also take part in the work. ISO collaborates closely with the
International Electrotechnical Commission (IEC) on all matters of electrotechnical standardization.
International Standards are drafted in accordance with the rules given in the ISO/IEC Directives, Part 2.
The main task of technical committees is to prepare International Standards. Draft International Standards
adopted by the technical committees are circulated to the member bodies for voting. Publication as an
International Standard requires approval by at least 75 % of the member bodies casting a vote.
Attention is drawn to the possibility that some of the elements of this document may be the subject of patent
rights. ISO shall not be held responsible for identifying any or all such patent rights.
ISO 80601-2-12 was prepared by Technical Committee ISO/TC 121, Anaesthetic and respiratory equipment,
Subcommittee SC 3, Lung ventilators and related equipment, and Technical Committee IEC/TC 62, Electrical
equipment in medical practice, Subcommittee SC D, Electrical equipment. The draft was circulated for voting
to the national bodies of both ISO and IEC.
This first edition of ISO 80601-2-12 cancels and replaces the second edition of IEC 60601-2-12:2001. This
edition of ISO 80601-2-12 constitutes a major technical revision of IEC 60601-2-12:2001 and includes an
alignment with the third edition of IEC 60601-1.
The most significant changes are the following modifications:
 extending the scope to include the critical care VENTILATOR and its ACCESSORIES, where the characteristics
of those ACCESSORIES can affect the BASIC SAFETY and ESSENTIAL PERFORMANCE of the VENTILATOR, and
thus not only the critical care VENTILATOR itself;
 identification of ESSENTIAL PERFORMANCE for a critical care VENTILATOR and its ACCESSORIES;
 modification of the obstruction of the expiratory limb (continuing AIRWAY PRESSURE) ALARM CONDITION
requirement;
and the following additions:
 tests for ventilation performance;
 tests for mechanical strength;
 new symbols;
 requirements for a critical care VENTILATOR as a component of an ME SYSTEM;
 tests for enclosure integrity (water ingress);
 tests for closed suction survivability of the VENTILATOR;
 tests for cleaning and disinfection procedures;
PATIENT from the gas pathways.
 consideration of contamination of the breathing gas delivered to the
vi © ISO 2011 – All rights reserved

---------------------- Page: 6 ----------------------
ISO 80601-2-12:2011(E)
ISO 80601 consists of the following parts, under the general title Medical electrical equipment:
 Part 2-12: Particular requirements for basic safety and essential performance of critical care ventilators
 Part 2-13: Particular requirements for basic safety and essential performance of an anaesthetic
workstation
 Part 2-55: Particular requirements for the basic safety and essential performance of respiratory gas
monitors
 Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for
body temperature measurement
 Part 2-61: Particular requirements for basic safety and essential performance of pulse oximeter
equipment for medical use
IEC 80601 consists of the following parts, under the general title Medical electrical equipment:
 IEC 80601-2-30: Particular requirements for the basic safety and essential performance of automated
non-invasive sphygmomanometers
 IEC 80601-2-35: Particular requirements for the basic safety and essential performance of heating
devices using blankets, pads and mattresses and intended for heating in medical use
 IEC 80601-2-58: Particular requirements for the basic safety and essential performance of lens removal
devices and vitrectomy devices for ophthalmic surgery
 IEC 80601-2-59: Particular requirements for the basic safety and essential performance of screening
thermographs for human febrile temperature screening
 IEC 80601-2-60: Particular requirements for basic safety and essential performance of dental equipment
The ISO and IEC 80601 family of standards are also parts of the IEC 60601 family of standards.

© ISO 2011 – All rights reserved vii

---------------------- Page: 7 ----------------------
ISO 80601-2-12:2011(E)
Introduction
In this International Standard, the following print types are used:
 Requirements and definitions: roman type.
 Test specifications: italic type.
 Informative material appearing outside of tables, such as notes, examples and references: in smaller
type. Normative text of tables is also in a smaller type.
 TERMS DEFINED IN IEC 60601-1:2005, CLAUSE 3, IN THIS PARTICULAR STANDARD OR AS NOTED: SMALL
CAPITALS TYPE.
In referring to the structure of this standard, the term
 “clause” means one of the seventeen numbered divisions within the table of contents, inclusive of all
subdivisions (e.g. Clause 7 includes subclauses 7.1, 7.2, etc.);
 “subclause” means a numbered subdivision of a clause (e.g. 201.7.1, 201.7.2 and 201.7.2.1 are all
subclauses of Clause 201.7).
References to clauses within this International Standard are preceded by the term “Clause” followed by the
clause number. References to subclauses within this particular International Standard are by number only.
In this International Standard, the conjunctive “or” is used as an “inclusive or” so a statement is true if any
combination of the conditions is true.
The verbal forms used in this International Standard conform to usage described in Annex H of the ISO/IEC
Directives, Part 2. For the purposes of this International Standard, the auxiliary verb:
 “shall” means that compliance with a requirement or a test is mandatory for compliance with this
standard;
 “should” means that compliance with a requirement or a test is recommended but is not mandatory for
compliance with this standard;
 “may” is used to describe a permissible way to achieve compliance with a requirement or test.
An asterisk (*) as the first character of a title or at the beginning of a paragraph or table title indicates that
there is guidance or rationale related to that item in Annex AA.
The attention of Member Bodies and National Committees is drawn to the fact that equipment manufacturers
and testing organizations may need a transitional period following publication of a new, amended or revised
ISO or IEC publication in which to make products in accordance with the new requirements and to equip
themselves for conducting new or revised tests. It is the recommendation of the committee that the content of
this International Standard not be adopted for mandatory implementation nationally earlier than 3 years from
the date of publication for equipment newly designed, and not earlier than 5 years from the date of publication
for equipment already in production.

viii © ISO 2011 – All rights reserved

---------------------- Page: 8 -------------------
...

NORME ISO
INTERNATIONALE 80601-2-12
Première édition
2011-04-15



Appareils électromédicaux —
Partie 2-12:
Exigences particulières relatives à la
sécurité de base et aux performances
essentielles des ventilateurs pulmonaires
pour utilisation en soins intensifs
Medical electrical equipment —
Part 2-12: Particular requirements for basic safety and essential
performance of critical care ventilators




Numéro de référence
ISO 80601-2-12:2011(F)
©
ISO 2011

---------------------- Page: 1 ----------------------
ISO 80601-2-12:2011(F)

DOCUMENT PROTÉGÉ PAR COPYRIGHT


©  ISO 2011
Droits de reproduction réservés. Sauf prescription différente, aucune partie de cette publication ne peut être reproduite ni utilisée sous
quelque forme que ce soit et par aucun procédé, électronique ou mécanique, y compris la photocopie et les microfilms, sans l'accord écrit
de l'ISO à l'adresse ci-après ou du comité membre de l'ISO dans le pays du demandeur.
ISO copyright office
Case postale 56  CH-1211 Geneva 20
Tel. + 41 22 749 01 11
Fax + 41 22 749 09 47
E-mail copyright@iso.org
Web www.iso.org
Publié en Suisse

ii © ISO 2011 – Tous droits réservés

---------------------- Page: 2 ----------------------
ISO 80601-2-12:2011(F)
Sommaire Page
Avant-propos . vii
Introduction . ix
201.1 Domaine d'application, objet et normes collatérales . 1
201.1. 1 Domaine d'application . 1
201.1. 2 Objet . 2
201.1. 3 Normes collatérales . 2
201.1. 4 Normes particulières . 2
201.2 Références normatives . 3
201.3 Termes et définitions . 6
201.4 Exigences générales . 9
201.4. 3 PERFORMANCES ESSENTIELLES. 9
201.4. 3.101 * Exigences supplémentaires relatives aux PERFORMANCES ESSENTIELLES . 9
201.4. 6 * Parties d'APPAREIL EM ou de SYSTÈME EM en contact avec le PATIENT . 10
201.4. 11.101 * Exigences supplémentaires relatives aux entrées de gaz sous pression . 10
201.4. 11.101.1 Exigences relatives à la surpression . 10
201.4. 11.101.2 Exigences de compatibilité . 10
201.5 Exigences générales d'essai d'un APPAREIL EM . 11
201.5. 101 * Exigences supplémentaires pour les exigences générales d'essai d'un
APPAREIL EM . 11
201.5. 101.1 Conditions d'essai du VENTILATEUR . 11
201.5. 101.2 * Spécifications des débits de gaz et des fuites . 11
201.5. 101.3 * Erreurs d'essai du VENTILATEUR . 11
201.6 Classification des APPAREILS EM et des SYSTÈMES EM . 11
201.7 Identification, marquage et documentation des APPAREILS EM . 11
201.7. 2.3 * Consultation de DOCUMENTS D'ACCOMPAGNEMENT . 12
201.7. 2.101 Exigences supplémentaires de marquage sur l'extérieur des APPAREILS EM ou
parties des APPAREILS EM . 12
201.7. 2.4.101 Exigences supplémentaires pour les ACCESSOIRES . 13
201.7. 2.13.101 Exigences supplémentaires en matière d'effets physiologiques . 13
201.7. 2.17.101 Exigences supplémentaires en matière d'emballage de protection . 13
201.7. 4.3 * Unité de mesure . 14
201.7. 9.1 Exigences générales supplémentaires . 14
201.7. 9.2.1.101 Exigences générales supplémentaires . 14
201.7. 9.2.2.101 * Exigences supplémentaires en matière d'avertissements et de consignes de
sécurité 14
201.7. 9.2.8.101 * Exigences supplémentaires en matière de procédure de démarrage . 15
201.7. 9.2.9.101 * Exigences supplémentaires en matière d'instructions de fonctionnement . 15
201.7. 9.2.12 Nettoyage, désinfection et stérilisation . 16
201.7. 9.2.14.101 * Exigences supplémentaires en matière d'ACCESSOIRES, d'équipements
supplémentaires et de fournitures utilisées . 16
201.7. 9.2.16.101 * Exigences supplémentaires en matière de référence à la description
technique 17
201.7. 9.3.1.101 * Exigences générales supplémentaires . 17
201.7. 9.3.101 Exigences supplémentaires en matière de description technique . 17
201.8 Protection contre les DANGERS d'origine électrique provenant des APPAREILS EM . 17
201.9 Protection contre les DANGERS mécaniques des APPAREILS EM et SYSTÈMES EM . 17
201.9. 6.2.1.101 Exigences supplémentaires en matière d'énergie acoustique . 18
© ISO 2011 – Tous droits réservés iii

---------------------- Page: 3 ----------------------
ISO 80601-2-12:2011(F)
201.9. 101 * Exigences supplémentaires en matière de procédures d'aspiration .19
201.10 Protection contre les DANGERS dus aux rayonnements involontaires ou excessifs .21
201.11 Protection contre les températures excessives et les autres DANGERS .21
201.11. 6.4 Fuite .21
201.11. 6.5.101 * Exigences supplémentaires en matière de pénétration d'eau ou de corps
solides dans l'APPAREIL EM ou le SYSTÈME EM .21
201.11. 6.6 * Nettoyage et désinfection des APPAREILS EM et des SYSTÈMES EM .22
201.11. 6.7 Stérilisation des APPAREILS EM ou des SYSTÈMES EM .22
201.11. 8.101 Exigences supplémentaires en matière de coupure de l'alimentation/du RÉSEAU
D'ALIMENTATION électrique vers l'APPAREIL EM .22
201.11. 8.101.1 CONDITION D'ALARME TECHNIQUE de panne d'alimentation électrique .22
201.11. 8.101.2 SOURCE ÉLECTRIQUE INTERNE ou source électrique en réserve externe .23
201.12 Précision des commandes, des instruments et de la protection contre les caractéristiques
de sortie présentant des risques .23
201.12. 1 * Précision des commandes et des instruments .23
201.12. 1.101 Ventilation en mode volume contrôlé .24
201.12. 1.102 Ventilation en mode pression contrôlée .27
201.12. 1.103 * SURVEILLANCE DU VOLUME DÉLIVRÉ .30
201.12. 1.104 * Réponse du VENTILATEUR à une augmentation de la teneur en O .30
2
201.12. 4 Protection contre les caractéristiques de sortie présentant des risques .32
201.12. 4.101 Moniteur d'oxygène .32
201.12. 4.102 * Mesure de la PRESSION DES VOIES AÉRIENNES .32
201.12. 4.103 * Mesure du volume expiré et CONDITIONS D'ALARME de faible volume .33
201.12. 4.103.1 VENTILATEURS destiné à fournir un VOLUME DÉLIVRÉ > 50 ml .33
201.12. 4.103.2 VENTILATEURS destinés à fournir un VOLUME DÉLIVRÉ ≤ 50 ml .33
201.12. 4.104 * DISPOSITIF DE PROTECTION DE LA PRESSION LIMITÉE MAXIMALE .34
201.12. 4.105 CONDITION D'ALARME de pression élevée et DISPOSITIF DE PROTECTION.34
201.12. 4.106 CONDITIONS D'ALARME DE PEEP .34
201.12. 4.107 * CONDITION D'ALARME d'obstruction .34
201.12. 4.108 * CONDITION D'ALARME d'occlusion partielle .35
201.12. 101 * Protection contre la modification accidentelle de commandes .35
201.13 SITUATIONS DANGEREUSES et conditions de défaut .35
201.13. 2.101 * CONDITIONS DE PREMIER DÉFAUT particulières supplémentaires .35
201.13. 102 * Défaillance d'une alimentation en gaz d'un VENTILATEUR .36
201.13. 103 * Indépendance de la fonction de commande de la ventilation et des mesures de
CONTRÔLE DU RISQUE associées .36
201.14 SYSTÈMES ÉLECTROMÉDICAUX PROGRAMMABLES (SEMP) .36
201.14. 101 Cycle de vie logiciel .36
201.15 Construction de l'APPAREIL EM .36
201.15. 3.5.101 Exigences supplémentaires en matière de manipulation brutale .36
201.15. 3.5.101.1 * Choc et vibrations .36
201.15. 3.5.101.2 * Choc et vibrations d'un VENTILATEUR MOBILE .37
201.15. 101 Mode de fonctionnement .38
201.15. 102 Teneur en oxygène délivré .39
201.15. 103 Contrôle automatique des ACCESSOIRES .39
201.16 SYSTÈMES EM .39
201.16. 1.101 Exigences générales supplémentaires en matière de SYSTÈMES EM.39
201.17 Compatibilité électromagnétique des APPAREILS EM et des SYSTÈMES EM .39
201.17. 101 Exigences supplémentaires en matière de compatibilité électromagnétique des
APPAREILS EM et des SYSTÈMES EM .39
201.101 Raccordements de gaz .39
201.101. 1 * Protection contre les rétropollutions .39
201.101. 2 Raccordement au système de distribution de gaz médicaux .40
201.101. 3 Raccords du VBS .40
201.101. 3.1 * Généralités .40
iv © ISO 2011 – Tous droits réservés

---------------------- Page: 4 ----------------------
ISO 80601-2-12:2011(F)
201.101. 3.2 Autres orifices désignés . 40
201.101. 3.2.1 Orifice de raccordement côté patient . 40
201.101. 3.2.2 Orifice de sortie du gaz et orifice de retour du gaz . 40
201.101. 3.2.3 Orifice de ventilation MANUELLE . 40
201.101. 3.2.4 Orifice d'aspiration de secours . 41
201.101. 3.2.5 Dispositifs conditionnant le sens du débit . 41
201.101. 3.2.6 Orifice d'accessoire . 41
201.101. 3.2.7 Orifice de la sonde de surveillance . 41
201.101. 3.2.8 Orifice d'évacuation du gaz . 41
201.102 Exigences relatives au VBS et à ses ACCESSOIRES . 41
201.102. 1 * Généralités . 41
201.102. 2 Étiquetage . 42
201.102. 3 Tubes respiratoires . 42
201.102. 4 * Gestion de l'eau . 42
201.102. 4.1 Système d'humidification . 42
201.102. 4.2 Échangeur de chaleur et d'humidité (ech) . 42
201.102. 5 Mélangeurs de gaz . 42
201.102. 6 Filtres de système respiratoire . 42
201.102. 7 Systèmes respiratoire du ventilateur . 42
201.102. 7.1 Fuite d'un VBS complet . 42
201.102. 7.2 * Ventilation non invasive . 43
201.103 * Respiration spontanée pendant la perte d'alimentation . 43
201.104 * Formation . 44
201.105 * Indication de la durée de fonctionnement . 44
201.106 ENTRÉE/SORTIE DE SIGNAL . 44
201.106. 1 Généralités . 44
201.106. 2 * Connexion à un dossier médical informatisé . 44
201.106. 3 * Connexion à un système d'alarme distribué . 44
201.106. 4 * Connexion à une commande à distance . 44
201.107 Affichage de courbes boucle . 45
201.107. 1 Boucles pression-volume . 45
201.107. 2 Boucles débit- volume . 45
201.108 * Pause ventilatoire temporisée . 45
201.108. 1 Pause expiratoire . 45
201.108. 2 Pause inspiratoire . 46
202 Appareils électromédicaux – Partie 1-2: Exigences générales pour la sécurité de base et
les performances essentielles – Norme collatérale: Compatibilité électromagnétique –
Exigences et essais . 46
202.6.2.1.10 * Critères de conformité . 46
206 Appareils électromédicaux – Partie 1-6: Exigences générales pour la sécurité de base et
les performances essentielles – Norme collatérale: Aptitude à l'utilisation . 47
208 Appareils électromédicaux – Partie 1-8: Exigences générales pour la sécurité de base et
les performances essentielles – Norme collatérale: Exigences générales, essais et guide
pour les systèmes d'alarme des appareils et des systèmes électromédicaux . 48
208.6.3.3.2.101 * Exigences supplémentaires en matière de caractéristiques de journalisation
de CONDITION D'ALARME . 48
208.6.8.3.101 Exigences supplémentaires en matière de désactivation globale indéfinie du
SIGNAL D'ALARME . 48
208.6.8.4.101 * Exigences supplémentaires en matière d'annulation de la désactivation des
SIGNAUX D'ALARME . 49
© ISO 2011 – Tous droits réservés v

---------------------- Page: 5 ----------------------
ISO 80601-2-12:2011(F)
Annexe C (informative) Guide des exigences en matière de marquage et d'étiquetage des
APPAREILS et SYSTÈMES EM .50
Annexe D (informative) Symboles des marquages .55
Annexe AA (informative) Lignes directrices particulières et justifications .57
Annexe BB (informative) Référence aux principes essentiels .76
Bibliographie .78
Index alphabétique des termes définis utilisés dans la présente norme particulière .81

vi © ISO 2011 – Tous droits réservés

---------------------- Page: 6 ----------------------
ISO 80601-2-12:2011(F)
Avant-propos
L'ISO (Organisation internationale de normalisation) est une fédération mondiale d'organismes nationaux de
normalisation (comités membres de l'ISO). L'élaboration des Normes internationales est en général confiée
aux comités techniques de l'ISO. Chaque comité membre intéressé par une étude a le droit de faire partie du
comité technique créé à cet effet. Les organisations internationales, gouvernementales et non
gouvernementales, en liaison avec l'ISO participent également aux travaux. L'ISO collabore étroitement avec
la Commission électrotechnique internationale (CEI) en ce qui concerne la normalisation électrotechnique.
Les Normes internationales sont rédigées conformément aux règles données dans les Directives ISO/CEI,
Partie 2.
La tâche principale des comités techniques est d'élaborer les Normes internationales. Les projets de Normes
internationales adoptés par les comités techniques sont soumis aux comités membres pour vote. Leur
publication comme Normes internationales requiert l'approbation de 75 % au moins des comités membres
votants.
L'attention est appelée sur le fait que certains des éléments du présent document peuvent faire l'objet de
droits de propriété intellectuelle ou de droits analogues. L'ISO ne saurait être tenue pour responsable de ne
pas avoir identifié de tels droits de propriété et averti de leur existence.
L'ISO 80601-2-12 a été élaborée par un groupe de travail mixte du comité technique ISO/TC 121, Matériel
d'anesthésie et de réanimation respiratoire, sous-comité SC 3, Ventilateurs pulmonaires et équipements
connexes et le Comité technique CEI/TC 62, Équipements électriques dans la pratique médicale, sous-comité
SC D, Appareils électromédicaux. Le projet a été circulé pour le vote aux membres nationaux de l'ISO et de la
CEI.
Cette première édition de l'ISO 80601-2-12 annule et remplace la deuxième édition de
la CEI 60601-2-12:2001. La présente édition de l'ISO 80601-2-12 constitue une révision technique majeure de
la CEI 60601-2-12:2001 et s'aligne sur la troisième édition de la CEI 60601-1.
Les modifications les plus significatives sont les suivantes:
 extension du domaine d'application pour inclure non seulement le VENTILATEUR en lui-même, mais
également ses ACCESSOIRES, lorsque les caractéristiques de ces ACCESSOIRES peuvent affecter la
SÉCURITÉ DE BASE et les PERFORMANCES ESSENTIELLES du VENTILATEUR;
 identification des PERFORMANCES ESSENTIELLES d'un VENTILATEUR pour utilisation en soins intensifs et ses
ACCESSOIRES;
 modification de l'exigence relative à la condition d'alarme d'obstruction de branche expiratoire (PRESSION
RESPIRATOIRE continue);
et les ajouts suivants:
 essais des performances de ventilation;
 essais de résistance mécanique;
 nouveaux symboles;
 exigences relatives au VENTILATEUR pour utilisation en soins intensifs en tant que composant d'un
SYSTÈME EM;
© ISO 2011 – Tous droits réservés vii

---------------------- Page: 7 ----------------------
ISO 80601-2-12:2011(F)
 essais d'intégrité de l'enveloppe (pénétration d'eau);
 essais de la capacité du VENTILATEUR à résister sans dommage à une procédure d'aspiration en circuit
fermé du VENTILATEUR;
 essais des modes opératoires de nettoyage et de désinfection; et
 considérations de contamination du gaz respiratoire délivré au PATIENT par les chemins du gaz.
L'ISO 80601 comprend les parties suivantes, présentées sous le titre général Appareils électromédicaux:
 Partie 2-12: Exigences particulières relatives à la sécurité de base et aux performances essentielles des
ventilateurs pulmonaires pour utilisation en soins intensifs
 Partie 2-13: Exigences particulières de sécurité de base et de performance essentielle pour les systèmes
d'anesthésie
 Partie 2-55: Exigences particulières relatives à la sécurité fondamentale et aux performances essentielles
des moniteurs de gaz respiratoire
 Partie 2-56: Exigences particulières relatives à la sécurité fondamentale et aux performances essentielles
des thermomètres médicaux pour mesurer la température de corps
 Partie 2-61: Exigences particulières pour la sécurité de base et les performances essentielles pour les
oxymètres de pouls
La CEI 80601 comprend les parties suivantes, présentées sous le titre général Appareils électromédicaux:
 CEI 80601-2-30: Exigences particulières de sécurité de base et de performances essentielles des
sphygmomanomètres non invasifs automatiques;
 CEI 80601-2-35: Exigences particulières pour la sécurité de base et les performances essentielles des
dispositifs de réchauffage utilisant des couvertures, des coussins ou des matelas chauffants et destinés
au réchauffage des patients en usage médical;
 CEI 80601-2-58: Exigences particulières pour la sécurité de base et les performances essentielles des
dispositifs de retrait du cristallin et des dispositifs de vitrectomie pour la chirurgie ophtalmique;
 CEI 80601-2-59: Exigences particulières relatives à la sécurité fondamentale et aux perfomances
essentielles des imageurs thermiques pour le dépistage des humains fébriles;
 CEI 80601-2-60: Exigences particulières pour la sécurité de base et les performances essentielles des
équipements dentaires.
La famille de normes de l'ISO et de la CEI 80601 font également partie de la famille de normes de la
CEI 60601.
viii © ISO 2011 – Tous droits réservés

---------------------- Page: 8 ----------------------
ISO 80601-2-12:2011(F)
Introduction
Dans la présente Norme internationale, les caractères suivants sont utilisés:
 exigences et définitions: caractères romains;
 modalités d'essai: caractères italiques.
 Les indications de nature informative apparaissant hors des tableaux, comme les notes, les exemples et
les références: petits caractères. Le texte normatif à l'intérieur des tableaux est également en petits
caractères.
 TERMES DÉFINIS DANS LA CEI 60601-1:2005, ARTICLE 3, DANS LA PRÉSENTE NORME OU EN NOTE: PETITES
LETTRES CAPITALES.
Concernant la structure de la présente norme, le terme
 «article» désigne l'une des dix-sept sections numérotées dans la table des matières, avec toutes ses
subdivisions (par exemple l'Article 7 inclut les paragraphes 7.1, 7.2, etc.);
 «paragraphe» désigne une subdivision numérotée d'un article (par exemple 201.7.1, 201.7.2 et 201.7.2.1
sont tous des paragraphes de l'Article 201.7).
Dans la présente Norme internationale, les références à des articles sont précédées du mot «Article» suivi du
numéro de l'article concerné. Dans la présente Norme internationale, les références aux paragraphes utilisent
uniquement le numéro du paragraphe concerné.
Dans la présente Norme internationale, la conjonction «ou» est utilisée avec la valeur d'un «ou inclusif», ainsi
un énoncé est vrai si une combinaison des conditions, quelle qu'elle soit, est vraie.
Les formes verbales utilisées dans la présente Norme internationale sont conformes à l'usage donné à
l'Annexe H des Directives ISO/CEI, Partie 2. Pour les besoins de la Norme internationale:
 «devoir» signifie que la conformité à une exigence ou à un essai est obligatoire pour la conformité à la
présente norme;
 «il convient/il est recommandé» signifie que la conformité à une exigence ou à un essai est
recommandée mais n'est pas obligatoire pour la conformité à la présente norme;
 «pouvoir» est utilisé pour décrire un moyen admissible pour satisfaire à une exigence ou un essai.
Lorsqu'un astérisque (*) est utilisé comme premier caractère devant un titre, au début d'un titre d'alinéa ou de
tableau, il indique l'existence d'une ligne directrice ou d'une justif
...