Respiratory protective devices - Compressed air line breathing devices with demand valve - Part 1: Devices with a full face mask - Requirements, testing and marking

This document specifies minimum requirements for compressed air line breathing devices with demand valve for use with a full face mask as a respiratory protective device (RPD). Escape and diving RPD and RPD used in abrasive blasting operations without additional protective features are not covered by this part of EN 14593, although certain requirements addressing the use in conjunction with escape RPD and escape conditions are given.
Laboratory and practical performance tests are included for the assessment of conformance to the requirements.

Atemschutzgeräte - Druckluft-Schlauchgeräte mit Lungenautomat - Teil 1: Geräte mit einer Vollmaske - Anforderungen, Prüfung und Kennzeichnung

Dieses Dokument legt Mindestanforderungen für Druckluft-Schlauchgeräte mit Lungenautomat und einer Vollmaske als Atemschutzgerät (ASG) fest. Für Fluchtgeräte und Tauchgeräte und ASG für Strahlarbeiten ohne zusätzliche Schutzeinrichtungen gilt dieser Teil der EN 14593 nicht, obwohl gewisse Anforderungen für den Gebrauch in Verbindung mit Fluchtgeräten und für Fluchtbedingungen angegeben werden.
Laborprüfungen und praktische Leistungsprüfungen sind enthalten, um die Übereinstimmung mit den Anforderungen feststellen zu können.

Appareils de protection respiratoire - Appareils de protection respiratoire isolants à adduction d'air comprimé avec soupape à la demande - Partie 1: Appareil avec masque complet - Exigences, essais et marquage

Le présent document spécifie les caractéristiques minimales exigées des appareils de protection respiratoire isolants à adduction d’air comprimé avec soupape à la demande et masque complet utilisés comme appareils de protection respiratoire (APR). Les APR destinés à l’évacuation, à la plongée et ceux utilisés dans les opérations de projection d’abrasifs sans dispositifs de protection supplémentaires ne sont pas couverts par la présente partie de l’EN 14593, bien que certaines exigences relatives à l’utilisation d’APR conjointement avec l’appareil d’évacuation et les conditions d’évacuation soient indiquées.
Il comporte des essais en laboratoire et des essais pratiques de performance pour évaluer la conformité aux exigences.

Oprema za varovanje dihal - Dihalni aparat na stisnjeni zrak z regulatorjem - 1. del: Aparat z obrazno masko - Zahteve, preskušanje in označevanje

Ta dokument določa minimalne zahteve za dihalne aparate na stisnjeni zrak z ventilom, ki so namenjeni za uporabo z obrazno masko in tvorijo opremo za varovanje dihal (RPD). Ta del standarda EN 14593 ne zajema opreme za varovanje dihal za izhod v sili ali potapljanje brez dodatnih zaščitnih funkcij, ki se uporablja pri peskanju, podaja pa določene zahteve glede uporabe skupaj z opremo za varovanje dihal za izhod v sili in pogoje za izhod v sili.
Vključeni so tudi laboratorijski in praktični preskusi učinkovitosti za ocenjevanje skladnosti z zahtevami.

General Information

Status
Published
Public Enquiry End Date
19-Feb-2017
Publication Date
06-Jun-2018
Current Stage
6060 - National Implementation/Publication (Adopted Project)
Start Date
29-May-2018
Due Date
03-Aug-2018
Completion Date
07-Jun-2018

Relations

Buy Standard

Standard
EN 14593-1:2018
English language
34 pages
sale 10% off
Preview
sale 10% off
Preview
e-Library read for
1 day

Standards Content (Sample)

2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.Atemschutzgeräte - Druckluft-Schlauchgeräte mit Lungenautomat - Teil 1: Geräte mit einer Vollmaske - Anforderungen, Prüfung und KennzeichnungAppareils de protection respiratoire - Appareils de protection respiratoire isolants à adduction d'air comprimé avec soupape à la demande - Partie 1: Appareil avec masque complet - Exigences, essais et marquageRespiratory protective devices - Compressed air line breathing devices with demand valve - Part 1: Devices with a full face mask - Requirements, testing and marking13.340.30Varovalne dihalne napraveRespiratory protective devicesICS:Ta slovenski standard je istoveten z:EN 14593-1:2018SIST EN 14593-1:2018en,fr,de01-julij-2018SIST EN 14593-1:2018SLOVENSKI
STANDARDSIST EN 14593-1:20051DGRPHãþD



SIST EN 14593-1:2018



EUROPEAN STANDARD NORME EUROPÉENNE EUROPÄISCHE NORM
EN 14593-1
May
t r s z ICS
s uä u v rä u r Supersedes EN
s v w { uæ sã t r r wEnglish Version
Respiratory protective devices æ Compressed air line breathing devices with demand valve æ Part
sã Devices with a full face mask æ Requirementsá testing and markingAppareils de protection respiratoire æ Appareils de protection respiratoire isolants à adduction d 5air comprimé avec soupape à la demande æ Partie
sã Appareil avec masque complet æ Exigencesá essais et marquage
Atemschutzgeräte æ DruckluftæSchlauchgeräte mit Lungenautomat æ Teil
sã Geräte mit einer Vollmaske æ Anforderungená Prüfung und Kennzeichnung This European Standard was approved by CEN on
s x March
t r s zä
egulations which stipulate the conditions for giving this European Standard the status of a national standard without any alterationä Upætoædate lists and bibliographical references concerning such national standards may be obtained on application to the CENæCENELEC Management Centre or to any CEN memberä
translation under the responsibility of a CEN member into its own language and notified to the CENæCENELEC Management Centre has the same status as the official versionsä
CEN members are the national standards bodies of Austriaá Belgiumá Bulgariaá Croatiaá Cyprusá Czech Republicá Denmarká Estoniaá Finlandá Former Yugoslav Republic of Macedoniaá Franceá Germanyá Greeceá Hungaryá Icelandá Irelandá Italyá Latviaá Lithuaniaá Luxembourgá Maltaá Netherlandsá Norwayá Polandá Portugalá Romaniaá Serbiaá Slovakiaá Sloveniaá Spainá Swedená Switzerlandá Turkey and United Kingdomä
EUROPEAN COMMITTEE FOR STANDARDIZATION COMITÉ EUROPÉEN DE NORMALISATION EUROPÄISCHES KOMITEE FÜR NORMUNG
CEN-CENELEC Management Centre:
Rue de la Science 23,
B-1040 Brussels
9
t r s z CEN All rights of exploitation in any form and by any means reserved worldwide for CEN national Membersä Refä Noä EN
s v w { uæ sã t r s z ESIST EN 14593-1:2018



EN 14593-1:2018 (E) 2 Contents Page European foreword . 5 1 Scope . 6 2 Normative references . 6 3 Terms, description and symbols . 7 3.1 Terms . 7 3.2 Description . 7 3.3 Symbols . 7 4 Requirements . 8 4.1 General . 8 4.2 Nominal values and tolerances . 8 4.3 Ergonomics . 8 4.4 Materials . 8 4.5 Practical performance . 9 4.6 Connections . 9 4.6.1 General . 9 4.6.2 Couplings . 9 4.6.3 Strength of connections to full face mask, demand valve, medium pressure connecting tube and breathing hose . 9 4.6.4 Connection between RPD and full face mask . 9 4.6.5 Unacceptable connections . 10 4.7 Compressed air line breathing device used with self-contained breathing device - switch over device . 10 4.7.1 General . 10 4.7.2 Switch over warning device. 10 4.8 Body harness or belt . 11 4.9 Pre-conditioning . 11 4.10 Flammability . 11 4.11 Resistance to pressure . 11 4.12 RPD with mobile high pressure air supply systems . 12 4.12.1 General . 12 4.12.2 Pressure reducer . 12 4.12.3 Pressure reducer relief valve . 12 4.13 Warning devices for mobile high pressure air supply systems. 12 4.13.1 General . 12 4.13.2 Audible warning device . 12 4.14 Compressed air supply tube . 13 4.14.1 Resistance to kinking . 13 4.14.2 Resistance to collapse . 13 4.14.3 Strength . 13 4.14.4 Flexibility . 13 4.14.5 Heat resistance . 13 4.14.6 Electrostatic properties . 13 4.14.7 Couplings . 14 4.15 Breathing hose . 14 4.15.1 Resistance to kinking . 14 4.15.2 Resistance to collapse . 14 SIST EN 14593-1:2018



EN 14593-1:2018 (E) 3 4.16 Lung governed demand valve . 14 4.16.1 Without positive pressure . 14 4.16.2 With positive pressure . 14 4.16.3 Supplementary air supply . 14 4.16.4 Couplings . 14 4.17 Adjustable parts . 15 4.18 Full face masks. 15 4.19 Total inward leakage . 15 4.20 Breathing resistance. 15 4.20.1 General . 15 4.20.2 Inhalation resistance . 15 4.20.3 Exhalation resistance . 15 4.21 Carbon dioxide content of inhalation air . 16 4.22 Leaktightness . 16 5 Testing . 16 5.1 General . 16 5.2 Inspection . 17 5.3 Practical performance . 18 5.3.1 General . 18 5.3.2 Preparation of RPD to be tested . 18 5.3.3 Test conditions . 18 5.3.4 Work simulation test . 19 5.3.5 Information to be recorded . 19 5.3.6 Practical performance tests at low temperature . 19 5.4 Strength of connections to facepiece, demand valve, medium pressure connecting tube and breathing hose . 19 5.5 Resistance to collapse of breathing hose . 20 5.5.1 Principle . 20 5.5.2 Apparatus . 20 5.5.3 Procedure . 20 5.6 Strength of compressed air supply tube, body harness and couplings . 21 5.7 Pre-conditioning . 22 5.8 Flammability . 22 5.9 Pressure relief valve . 22 5.10 Resistance to kinking of compressed air supply tube . 22 5.11 Resistance to collapse of compressed air supply tube . 25 5.11.1 Principle . 25 5.11.2 Apparatus . 25 5.11.3 Procedure . 25 5.12 Heat resistance of compressed air supply tube . 25 5.13 Tests for lung-governed demand valve . 26 5.13.1 Without positive pressure . 26 5.13.2 With positive pressure . 26 5.13.3 Supplementary air supply . 26 5.14 Determination of carbon dioxide content of the inhalation air. 27 5.15 Testing of audible warning device . 27 5.16 Switch over device . 28 5.17 Leaktightness . 28 5.18 Sound level measurement . 28 6 Marking . 29 7 Information supplied by the manufacturer . 29 SIST EN 14593-1:2018



EN 14593-1:2018 (E) 4 Annex ZA (informative)
Relationship between this European Standard and the essential health and safety requirements of Regulation 2016/425/EU [2016 OJ L81] aimed to be covered . 32 Bibliography . 34
SIST EN 14593-1:2018



EN 14593-1:2018 (E) 5 European foreword This document (EN 14593-1:2018) has been prepared by Technical Committee CEN/TC 79 “Respiratory protective devices”, the secretariat of which is held by DIN. This European Standard shall be given the status of a national standard, either by publication of an identical text or by endorsement, at the latest by November 2018, and conflicting national standards shall be withdrawn at the latest by November 2018. Attention is drawn to the possibility that some of the elements of this document may be the subject of patent rights. CEN not be held responsible for identifying any or all such patent rights. This document supersedes EN 14593-1:2005. This document has been prepared under a standardization request given to CEN by the European Commission and the European Free Trade Association, and supports essential health and safety requirements of EU Regulation(s). For relationship with EU Regulation(s) see informative Annex ZA, which is an integral part of this document. The following main technical changes have been made compared to EN 14593-1:2005: a) requirements for cleaning and disinfection deleted; b) visual inspection changed to inspection and detailed list inserted; c) test for leaktightness added; d) test for noise level adapted to the test procedure specified in ISO 16900-14; e) Annex B deleted; f) figures adapted to the changes made in the test procedures, where appropriate. According to the CEN/CENELEC Internal Regulations, the national standards organizations of the following countries are bound to implement this European Standard: Austria, Belgium, Bulgaria, Croatia, Cyprus, Czech Republic, Denmark, Estonia, Finland, Former Yugoslav Republic of Macedonia, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta, Netherlands, Norway, Poland, Portugal, Romania, Serbia, Slovakia, Slovenia, Spain, Sweden, Switzerland, Turkey and the United Kingdom. SIST EN 14593-1:2018



EN 14593-1:2018 (E) 6 1 Scope This document specifies minimum requirements for compressed air line breathing devices with demand valve for use with a full face mask as a respiratory protective device (RPD). Diving RPD are not covered by this document. RPD used in abrasive blasting operations without additional protective features are not covered by this document. Escape RPD, although certain requirements addressing the use in conjunction with escape RPD and escape conditions are given, are not covered by this document. Laboratory and practical performance tests are included for the assessment of conformance to the requirements. 2 Normative references The following documents are referred to in the text in such a way that some or all of their content constitutes requirements of this document. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies. EN 132:1998, Respiratory protective devices — Definitions of terms and pictograms EN 134:1998, Respiratory protective devices — Nomenclature of components EN 136:1998, Respiratory protective devices — Full face masks — Requirements, testing, marking EN 137:2006, Respiratory protective devices — Self-contained open-circuit compressed air breathing apparatus with full face mask — Requirements, testing, marking EN 148-1, Respiratory protective devices — Threads for facepieces — Part 1: Standard thread connection EN 148-2, Respiratory protective devices — Threads for facepieces — Part 2: Centre thread connection EN 148-3, Respiratory protective devices — Threads for facepieces — Part 3: Tread connection M 45 x 3 EN 402:2003, Respiratory protective devices — Lung governed demand self-contained open-circuit compressed air breathing apparatus with full face mask or mouthpiece assembly for escape — Requirements, testing, marking EN 12021, Respiratory equipment — Compressed gases for breathing apparatus EN 13274-1:2001, Respiratory protective devices — Methods of test — Part 1: Determination of inward leakage and total inward leakage EN 13274-2:2001, Respiratory protective devices — Methods of test — Part 2: Practical performance tests EN 13274-3:2001, Respiratory protective devices — Methods of test — Part 3: Determination of breathing resistance EN 13274-4:2001, Respiratory protective devices — Methods of test — Part 4: Flame tests EN 13274-6, Respiratory protective devices — Methods of test — Part 6: Determination of carbon dioxide content of the inhalation air SIST EN 14593-1:2018



EN 14593-1:2018 (E) 7 EN ISO 8031, Rubber and plastics hoses and hose assemblies — Determination of electrical resistance and conductivity (ISO 8031) ISO 16900-14, Respiratory protective devices — Methods of test and test equipment — Part 14: Measurement of sound level 3 Terms, description and symbols For the purposes of this document, the terms, definitions and symbols given in EN 132:1998, EN 134:1998 and the following apply. 3.1 Terms 3.1.1 as received not pre-conditioned or modified to carry out a test 3.1.2 facepiece [Respiratory Interface (RI)] full face mask conforming to EN 136 (Class 2 or Class 3) 3.1.3 mobile high pressure air supply system supply system that can include a compressor, filters, compressed air cylinders, for use as a mobile source of breathable air 3.1.4 switch over device device that ensures that in the event of a malfunction or disconnection of the airline, the air supply, when airline is used in conjunction with self-contained breathing devices, automatically switches over to the self contained air supply, without interruption of supplied air to the wearer 3.2 Description This RPD supplies the wearer with breathable air in accordance with EN 12021, which on inhalation, flows through a lung governed demand valve. The latter can operate at negative pressure or at positive pressure and is connected to a suitable full face mask, possibly via a breathing hose. A compressed air supply tube connects the wearer to a supply of compressed air. Exhaled air flows into the ambient atmosphere via an exhalation valve. NOTE Conformance to EN 12021 can be ensured by a breathable air supply system or an additional device such as a compressed air filter system. 3.3 Symbols 3.3.1
See information supplied by the RPD manufacturer SIST EN 14593-1:2018



EN 14593-1:2018 (E) 8 4 Requirements 4.1 General All test samples specified in the related test clauses shall meet the relevant requirements. Where it is required in a specific clause, the manufacturer shall declare that a risk assessment e.g. a Failure Modes and Effect Analysis (FMEA) concerning these specific requirements has been conducted. NOTE Further information is given in EN 60812 [3]. If the manufacturer claims that the RPD may be used in potentially explosive atmospheres it shall comply with the appropriate requirements. 4.2 Nominal values and tolerances Unless otherwise specified, the values stated in this document are expressed as nominal values. Except for temperature limits, values which are not stated as maxima or minima shall be subject to a tolerance of ± 5 %. Unless otherwise specified, the ambient conditions for testing shall be between 16 °C and 32 °C and (50 ± 30) % relative humidity. Any temperature limits specified shall be subject to an accuracy of ± 1°C. 4.3 Ergonomics The requirements of this document are intended to take account of the interaction between the wearer, the RPD, and where possible the working environment in which the RPD is likely to be used. The RPD shall satisfy 4.4 and 4.8 . Testing shall be performed accordance with 5.3. 4.4 Materials 4.4.1 All materials used in the construction shall have adequate resistance to deterioration by heat and adequate mechanical strength. After any pre-conditioning according to 5.7, testing shall be performed in accordance with 5.3. The manufacturer shall supply a declaration that this was addressed by a risk assessment, e.g. a FMEA. Check in accordance with 5.2. 4.4.2 If the manufacturer claims that the RPD may be used in potentially explosive atmospheres, exposed parts, i.e. those which can be subjected to impact during use of the RPD shall not be made of aluminium, magnesium, titanium or their alloys. The manufacturer shall supply a declaration that this was addressed by a risk assessment, e.g. a FMEA. Check in accordance with 5.2. 4.4.3 Materials that can come into direct contact with the wearer's skin or that can affect the quality of the breathed air shall not be known to be likely to cause skin irritation or any other adverse effects to health. The manufacturer shall supply a declaration that this was addressed by a risk assessment, e.g a FMEA. Check in accordance with 5.2. 4.4.4 The finish of any part of the RPD likely to be in contact with the wearer shall be free from sharp edges and burrs. Check in accordance with 5.2 and test in accordance with 5.3. SIST EN 14593-1:2018



EN 14593-1:2018 (E) 9 4.5 Practical performance The complete RPD shall undergo practical performance tests under realistic conditions. These general tests serve the purpose of checking the RPD for imperfections that cannot be determined by the tests specified elsewhere in this document. If during any activity, by any test subject, the test subject fails to finalise the selected activity due to the RPD being not fit for the purpose for which it has been designed, the RPD shall be deemed to have failed. After completion of the activities the test subjects are asked to answer the questions in EN 13274-2:2001, 6.6. Testing shall be performed in accordance with 5.3. 4.6 Connections 4.6.1 General Components of the RPD shall be readily separated for cleaning, examining and testing. All demountable connections shall be readily connected and secured, where possible by hand. Any means of sealing used shall be retained in position when the joints and couplings are disconnected during normal use and maintenance. Check in accordance with 5.2 and test in accordance with 5.3. 4.6.2 Couplings The RPD shall be constructed so that any twisting of the hoses and tubes does not affect the fit or performance of the RPD, or cause the hoses or tubes to become disconnected. At least one swivelling coupling shall be fitted to the compressed air supply tube adjacent to the wearer. The design of the couplings shall be such as to prevent unintentional interruption of the air supply. Check in accordance with 5.2 and test in accordance with 5.3. 4.6.3 Strength of connections to full face mask, demand valve, medium pressure connecting tube and breathing hose Connections of the breathing hose at the full face mask and at the demand valve or between the full face mask connector and the demand valve shall withstand a force of 250 N. Testing shall be performed in accordance with 5.4. 4.6.4 Connection between RPD and full face mask The connection between the RPD and the full face mask can be achieved by a permanent, special or thread type connector. If a thread connector is used, either it shall conform to the requirements of one of the following standards: — EN 148-1, for RPD without positive pressure, — EN 148-3, for RPD with posit
...