IEC 61557-16:2023 specifies the requirements applicable to the performance for test and measurement equipment in order to determine the effectiveness of the protective measures for electrical equipment and/or medical electrical equipment described in IEC 62353.
This International Standard is to be used in conjunction with IEC 61557-1:2019.
IEC 61557-16:2023 cancels and replaces the first edition published in 2014. This edition constitutes a technical revision.
This edition includes the following significant technical changes with respect to the previous edition:
a) splitting of uncertainty requirements for medical and non-medical electrical equipment in 4.2.1;
b) addition of a definition of ranges with defined uncertainty in 4.2.1 to 4.2.7;
c) addition of an optional measuring device (MD) for non-medical devices in 4.2.1;
d) addition of a limitation of the maximum intrinsic uncertainty for medical applications at leakage current in 4.2.1;
e) change of 4.2.3 from test sockets to sockets for service purposes;
f) addition of a warning in the operating instructions;
g) integration of former 6.3 into 6.2;
h) update of Table 1;
i) alignment of the structure with that of the whole IEC 61557 series.

  • Standard
    42 pages
    English and French language
    sale 15% off

IEC 61557-14:2023 defines special requirements for test and measurement equipment used to determine the electrical safety of electrical equipment of machinery in accordance with IEC 60204-1.
This International Standard is to be used in conjunction with IEC 61557-1:2019.
IEC 61557-14:2023 cancels and replaces the first edition published in 2013. This edition constitutes a technical revision.
This edition includes the following significant technical changes with respect to the previous edition:
a) clarifying the introduction;
b) replaced "dielectric strength" by "voltage test";
c) requirement for maximum output current has been added in 4.2.6.1;
d) tripping time at electrical switching activated by two-hand operation has been added in 4.2.6.1;
e) additional time limiting capability for the protection against electric shock for test persons and bystanders in 4.2.6.2;
f) updated references for safety testing;
g) alignment of the structure with that of the whole IEC 61557 series.

  • Standard
    33 pages
    English and French language
    sale 15% off

IEC 61557-7:2019 is available as IEC 61557-7:2019 RLV which contains the International Standard and its Redline version, showing all changes of the technical content compared to the previous edition.IEC 61557-7:2019 specifies the requirements applicable to measuring equipment for testing the phase sequence in three-phase distribution systems. Indication of the phase sequence can be mechanical, visual and/or audible. This document does not apply to additional measurements for other quantities. It does not apply to monitoring relays. IEC 61557-7:2019 cancels and replaces the second edition published in 2007. This edition constitutes a technical revision. This edition includes the following changes with respect to the previous edition:
a) alignment of the structure with that of the whole IEC 61557 series;
b) updated requirements in 4.3 in accordance with new editions of IEC 61010-1 and IEC 61010-031;
c) the information on markings was extended;
d) the information on the operating instructions was extended;
e) complement to the information on the testing of leads;
f) test leads for insulated conductors were introduced;
g) Annex B was added with information on phase sequence tests and indications.

  • Standard
    75 pages
    English and French language
    sale 15% off
  • Standard
    28 pages
    English and French language
    sale 15% off

IEC 61557-13:2023 defines special performance requirements for hand-held and hand manipulated current clamps and sensors for measurement of leakage currents in electrical distribution systems up to 1 000 V AC and 1 500 V DC taking into account the influence of high external low-frequency magnetic fields and other influencing quantities. See Annex A for examples of measurement applications.
This document does not apply to current clamps or sensors that are used in combination with devices for insulation fault location in accordance with IEC 61557-9, unless it is specified by the manufacturer.
IEC 61557-13:2023 cancels and replaces the first edition published in 2011. This edition constitutes a technical revision.
This edition includes the following significant technical changes with respect to the previous edition:
a) the term "fixing device" has been removed;
b) the measuring range was changed to a display range, the indication of DC or peak values has been added in 4.1;
c) the frequency for the test of sensitivity for low-frequency magnetic fields has been defined in 4.2;
d) the specified measuring range is now defined as the range of indicated values based on the operating uncertainty in 4.3;
e) alignment of the structure with that of the whole IEC 61557 series;
f) the variation E12 (maximum load current), may be specified according to the manufacturer’s specification.

  • Standard
    41 pages
    English and French language
    sale 15% off

IEC 61557-17:2021 defines minimum performance requirements for non-contact AC voltage indicators to reduce the risk of electric shock for the testing person and bystanders caused by the wrong interpretation of the indication.
Products designed and manufactured in accordance with this document are for use by (electrically) skilled persons only. Non-contact AC voltage indicators are not designed for testing the absence of the operating voltage.
This International Standard is to be used in conjunction with IEC 61557-1:2019.

  • Standard
    25 pages
    English and French language
    sale 15% off

IEC 61557-11:2020 is available as IEC 61557-11:2020 RLV which contains the International Standard and its Redline version, showing all changes of the technical content compared to the previous edition.
IEC 61557-11:2020 specifies the requirements for test equipment applied to the testing of the effectiveness of residual current monitors (RCM) that are already installed in distribution systems. This test equipment can be used in any kind of network, such as a TN, TT or IT system. The test equipment can also be used for testing directionally discriminating residual current monitors (RCM) in IT systems. It is not the purpose of this document to verify the residual current monitors (RCM) according to their product standards. IEC 61557-11:2020 cancels and replaces the first edition published in 2009. This edition constitutes a technical revision. This edition includes the following significant technical changes with respect to the previous edition:
a) document title modified to include all types of RCM;
b) terms aligned with IEC 60050;
c) addition of requirements for testing new types of RCM;
d) moving of requirements for RCM Type B from former Annex A to main body text;
e) alignment of the structure with that of the whole IEC 61557 series.

  • Standard
    29 pages
    English and French language
    sale 15% off

IEC 61557-1:2019 is available as IEC 61557-1:2019 RLV which contains the International Standard and its Redline version, showing all changes of the technical content compared to the previous edition.
IEC 61557-1:2019 specifies the general requirements applicable to measuring and monitoring equipment for testing the electrical safety in low-voltage distribution systems with nominal voltages up to 1 000 V AC and 1 500 V DC. When measuring equipment or measuring installations involve measurement tasks of various measuring equipment covered by this series of standards, then the part of this series relevant to each of the measurement tasks is applicable. Other parts of IEC 61557 can specify additional requirements or deviations. This document does not cover functional safety or cybersecurity. IEC 61557-1:2019 cancels and replaces the second edition published in 2007. This edition constitutes a technical revision. This third edition includes the following significant technical changes with respect to the previous edition:
a) terms aligned with IEC 60050;
b) measurement of uncertainty revised according to the equations in 4.2 of ISO/IEC Guide 98-3:2008 (GUM);
c) updated references for safety and EMC requirements;
d) updated references for marking and operating instructions;
e) updated references for testing safety and EMC;
f) Annex A contains an explanation of GUM;
g) Annex B addresses environmental aspects.

  • Standard
    50 pages
    English and French language
    sale 15% off

IEC 61557-6:2019 is available as IEC 61557-6:2019 RLV which contains the International Standard and its Redline version, showing all changes of the technical content compared to the previous edition.IEC 61557-6:2019 specifies the requirements applicable to measuring equipment for testing the effectiveness of protective measures of residual current devices (RCD) installed in TT, TN and IT systems. It is not the purpose of this document to verify the RCD according to their product standards. IEC 61557-6:2019 cancels and replaces the second edition published in 2007. This edition constitutes a technical revision. This edition includes the following significant technical changes with respect to the previous edition:
a) addition of requirements for testing a new type of RCD;
b) addition of requirements for type B RCDs (former Annex B);
c) addition of new Annex B on recommended tripping times;
d) alignment of the structure with that of the whole IEC 61557 series.

  • Standard
    28 pages
    English and French language
    sale 15% off

IEC 61557-2:2019 is available as IEC 61557-2:2019 RLV which contains the International Standard and its Redline version, showing all changes of the technical content compared to the previous edition.IEC 61557-2:2019 specifies the requirements applicable to equipment for measuring the insulation resistance of equipment and installations in the de-energized state IEC 61557-2:2019 cancels and replaces the second edition published in 2007. This edition constitutes a technical revision. This edition includes the following significant technical changes with respect to the previous edition:
a) addition of requirements as regards measurement category;
b) addition of new requirements for operating instructions;
c) alignment of the structure with that of the the whole IEC 61557 series.

  • Standard
    19 pages
    English and French language
    sale 15% off

IEC 61557-5:2019 is available as IEC 61557-5:2019 RLV which contains the International Standard and its Redline version, showing all changes of the technical content compared to the previous edition.IEC 61557-5:2019 specifies the requirements applicable to measuring equipment for measuring the resistance to earth using an AC voltage. IEC 61557-5:2019 cancels and replaces the second edition published in 2007. This edition constitutes a technical revision. This edition includes the following significant technical changes with respect to the previous edition:
a) definitions and symbols in Clause 3 modified;
b) subclauses in Clause 4 restructured and aligned with other parts of the series;
c) limits for reduced voltages 25 V RMS or 35 V peak removed from 4.5;
d) requirements for clamps added;
e) marking for rated voltages to earth and measurement category added to Clause 5;
f) warning about absence of hazardous voltage added in Clause 5;
g) the term "percentage operating uncertainty" replaced by "operating uncertainty" in Clause 6;
h) equation for uncertainty corrected in Table 1;
i) new Annex A on test measurements with loop clamps added.

  • Standard
    24 pages
    English and French language
    sale 15% off

IEC 61557-3:2019 is available as IEC 61557-3:2019 RLV which contains the International Standard and its Redline version, showing all changes of the technical content compared to the previous edition.IEC 61557-3:2019 specifies the requirements applicable to equipment for measuring the loop impedance between a line conductor and protective conductor; between a line conductor and neutral; or between two line conductors by using the voltage drop when the circuit under test is loaded. IEC 61557-3:2019 cancels and replaces the second edition published in 2007. This edition constitutes a technical revision. This edition includes the following significant technical changes with respect to the previous edition:
a) addition of equirements as regards the measurement category;
b) addition of new requirements for operating instructions;
c) alignment of the structure with that of the whole IEC 61557 series.

  • Standard
    21 pages
    English and French language
    sale 15% off

IEC 61557-4:2019 is available as IEC 61557-4:2019 RLV which contains the International Standard and its Redline version, showing all changes of the technical content compared to the previous edition.IEC 61557-4:2019 specifies the requirements applicable to equipment for measuring the resistance of earth conductors, protective earth conductors and conductors for equipotential bonding, including their connections and terminals, with an indication of the measured value or an indication of the limits. IEC 61557-4:2019 cancels and replaces the second edition published in 2007. This edition constitutes a technical revision. This edition includes the following significant technical changes with respect to the previous edition:
a) complement to the measurement category in Clause 4;
b) correction of the equation for operating uncertainty;
c) complement to the requirements for measuring with DC;
d) alignment of the structure with that of the whole IEC 61557 series.

  • Standard
    20 pages
    English and French language
    sale 15% off

IEC 61557-8:2014 specifies the requirements for insulation monitoring devices (IMD) which permanently monitor the insulation resistance RF to earth of unearthed a.c. IT systems, of a.c. IT systems with galvanically connected d.c. circuits having nominal voltages up to 1 000 V a.c., as well as of unearthed d.c. IT systems with voltages up to 1 500 V d.c. independent from the method of measuring. IT systems are described in IEC 60364-4-41 amongst other literature. Additional data for the selection of devices in other standards should be noted. IMDs according to this part of IEC 61557 can also be used for de-energized TT, TN and IT systems or appliances. This third edition cancels and replaces the second edition published in 2007. This edition constitutes a technical revision which includes the following significant technical changes with respect to the previous edition:
- terms and definitions have been complemented;
- abbreviations are listed and explained;
- requirements have been revised;
- mandatory and optional functions and their terminology have been adapted from IEC 61557-15;
- mechanical requirements have been added;
- information on operating instructions has been added;
- type tests and routine tests have been complemented;
- an Annex C: 'Insulation monitoring devices for photovoltaic systems (PV-IMD)' has been added; and
- an Annex D: 'Insulation monitoring function of a photovoltaic inverter (PV-IMF) or in a charge controller' has been added.
This publication is to be read in conjunction with IEC 61557-1:2007. The contents of the corrigendum of May 2016 have been included in this copy.

  • Standard
    95 pages
    English and French language
    sale 15% off

IEC 61557-16:2014 defines performance requirements for test and measurement equipment to determine the effectiveness of the protective measures of electrical measures for electrical equipment and/or medical electrical equipment described in IEC 62353. It is the intention of this standard to achieve comparable measuring results, additional safety for the testing person and negligible electrical stress for the unit under test. Keywords: leakage current, patient applied part
This publication is to be read in conjunction with IEC 61557-1:2007.

  • Standard
    38 pages
    English and French language
    sale 15% off

IEC 61557-9:2014 specifies the requirements for the insulation fault location system (IFLS) which localizes insulation faults in any part of the system in unearthed IT a.c. systems and unearthed IT a.c. systems with galvanically connected d.c. circuits having nominal voltages up to 1 000 V a.c., as well as in unearthed IT d.c. systems with voltages up to 1 500 V d.c., independent of the measuring principle. IT systems are described in IEC 60364-4-41 amongst other literature. Additional data for a selection of devices in other standards should be noted. This third edition cancels and replaces the second edition published in 2009. This edition constitutes a technical revision which includes the following significant technical changes with respect to the previous edition:
- the scope, normative references, terms and definitions have been complemented;
- abbreviations are listed and explained;
- requirements, marking and operating instructions have been revised;
- mandatory and optional functions have been defined and their terminology has been adapted to IEC 61557-15;
- mechanical requirements have been added;
- Clause 6 "Tests" has been revised; and
- new Tables have been added.
This publication is to be read in conjunction with IEC 61557-1:2007. The contents of the corrigenda of May 2016 and of January 2017 have been included in this copy.

  • Standard
    55 pages
    English and French language
    sale 15% off

IEC 61557-15:2014 specifies requirements related to functional safety and is based on the IEC 61508 standard series for the realization of Insulation Monitoring Devices (IMD) as specified in IEC 61557-8 and for Insulation Fault Location Systems (IFLS) according to IEC 61557-9, according to phase 10 of the IEC 61508-1 lifecycle. These devices provide safety related functions for IT systems. This part of IEC 61557 is
- concerned only with functional safety requirements intended to reduce the functional risk during the use of IMDs and IFLSs;
- restricted to risks arising directly from the device itself or from several IMDs or IFLSs working together in a system; and
- intended to define the basic safety functions provided by the devices. This part of IEC 61557 does not
- deal with electrical safety according to IEC 61010-1 and the requirements of IEC 61557-8 and IEC 61557-9;
- cover the hazard and risk analysis of a particular use of the IMD or IFLS;
- identify all the safety functions for the application in which the IMD or IFLS is used; and
- cover the IMD or IFLS manufacturing process. Functional safety requirements depend on the application and should be considered as part of the overall risk assessment of the specific application. The supplier of IMDs and IFLSs is not responsible for the application. The application designer is responsible for the risk assessment and for specifying the overall functional safety requirements of the complete IT system and he should select the functional safety level (SIL) of the IMD and/or IFLS when their safety function is part of the functional safety assessment in the IT system.
This publication is to be read in conjunction with IEC 61557-8:2007 and IEC 61557-9:2009

  • Standard
    154 pages
    English and French language
    sale 15% off

IEC 61557-10:2013 specifies the requirements for combined measuring equipment which combines into one piece of apparatus, several measuring functions or methods of testing, measuring or monitoring according to the respective parts of IEC 61557. This second edition cancels and replaces the first edition published in 2000. This edition constitutes a technical revision. (electrical safety, combined measuring equipment for testing, measuring or monitoring of protective measures)
This publication is to be read in conjunction with IEC 61557-1:2007.

  • Standard
    17 pages
    English and French language
    sale 15% off

IEC 61557-14:2013 defines special requirements for test and measurement equipment used to determine the electrical safety of electrical equipment of machinery according to IEC 60204-1. IEC 61010 and the existing parts of series IEC 61557 do not cover all safety aspects of testing electrical equipment of machinery. IEC 61557-14 provides additional protection against electric shock for the testing person and bystanders during high-voltage-tests and in case of unintended use of the test equipment. It defines performance requirements for each measuring and testing function to ensure comparable results.

  • Standard
    33 pages
    English and French language
    sale 15% off

IEC 61557-13:2011 defines special performance requirements for hand-held and hand-manipulated current clamps and sensors for measurement of leakage currents in electrical distribution systems up to 1 000 V a.c. and 1 500 V d.c., taking into account the influence of high external low-frequency magnetic fields and other influencing quantities. This standard does not apply to current clamps or sensors which are used in combination with devices for insulation fault location according to IEC 61557-9, unless it is specified by the manufacturer.

  • Standard
    37 pages
    English and French language
    sale 15% off

IEC 61557-9:2009 specifies the requirements for insulation fault location systems which localize insulation faults in any part of the system in unearthed IT a.c. systems and unearthed IT a.c. systems with galvanically connected d.c. circuits having nominal voltages up to 1 000 V a.c., as well as in unearthed IT d.c. systems with voltages up to 1 500 V d.c., independent of the measuring principle. This second edition cancels and replaces the first edition published in 1999. It is a technical revision. This edition includes the following significant technical changes with respect to the previous edition:
- scope, normative references and terms and definitions complemented;
- revision of requirements, marking and operating instructions and Clause 6;
- revision of Table 1;
- addition of Annexes A, B and C.

  • Standard
    45 pages
    English and French language
    sale 15% off

IEC 61557-11:2009 specifies the requirements for testing equipment applied to the testing of the effectiveness of residual current monitors (RCMs) of type A and type B, which are already installed in distribution systems. This test equipment can be used in any kind of network like a TN, TT or IT system. The test equipment may also be used for testing directionally discriminating RCMs in IT-Systems.This part is to be used in conjunction with IEC 61557-1:2007, Part 1: General requirements.

  • Standard
    41 pages
    English and French language
    sale 15% off